Your browser doesn't support javascript.
loading
Equilibrium phase contrast-enhanced magnetic resonance angiography of the thoracic aorta and heart using balanced T1 relaxation-enhanced steady-state.
Edelman, Robert R; Ozturk, Onural; Pursnani, Amit; Balasubramanian, Senthil; Leloudas, Nondas; Koktzoglou, Ioannis.
Afiliação
  • Edelman RR; Department of Radiology, Northshore University HealthSystem, Evanston, Illinois, USA; Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA. Electronic address: redelman999@gmail.com.
  • Ozturk O; Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA.
  • Pursnani A; Department of Radiology, Northshore University HealthSystem, Evanston, Illinois, USA; Department of Medicine, Pritzker School of Medicine, University of Chicago, Chicago, Illinois, USA.
  • Balasubramanian S; Department of Radiology, Northshore University HealthSystem, Evanston, Illinois, USA.
  • Leloudas N; Department of Radiology, Northshore University HealthSystem, Evanston, Illinois, USA.
  • Koktzoglou I; Department of Radiology, Northshore University HealthSystem, Evanston, Illinois, USA; Department of Radiology, Pritzker School of Medicine, University of Chicago, Chicago, Illinois, USA.
J Cardiovasc Magn Reson ; 26(2): 101046, 2024 May 27.
Article em En | MEDLINE | ID: mdl-38810732
ABSTRACT

BACKGROUND:

Three-dimensional (3D) contrast-enhanced magnetic resonance angiography (CEMRA) is routinely used for vascular evaluation. With existing techniques for CEMRA, diagnostic image quality is only obtained during the first pass of the contrast agent or shortly thereafter, whereas angiographic quality tends to be poor when imaging is delayed to the equilibrium phase. We hypothesized that prolonged blood pool contrast enhancement could be obtained by imaging with a balanced T1 relaxation-enhanced steady-state (bT1RESS) pulse sequence, which combines 3D balanced steady-state free precession (bSSFP) with a saturation recovery magnetization preparation to impart T1 weighting and suppress background tissues. An electrocardiographic-gated, two-dimensional-accelerated version with isotropic 1.1-mm spatial resolution was evaluated for breath-hold equilibrium phase CEMRA of the thoracic aorta and heart.

METHODS:

The study was approved by the institutional review board. Twenty-one subjects were imaged using unenhanced 3D bSSFP, time-resolved CEMRA, first-pass gated CEMRA, followed by early and late equilibrium phase gated CEMRA and bT1RESS. Nine additional subjects were imaged using equilibrium phase 3D bSSFP and bT1RESS. Images were evaluated for image quality, aortic root sharpness, and visualization of the coronary artery origins, as well as using standard quantitative measures.

RESULTS:

Equilibrium phase bT1RESS provided better image quality, aortic root sharpness, and coronary artery origin visualization than gated CEMRA (P < 0.05), and improved image quality and aortic root sharpness versus unenhanced 3D bSSFP (P < 0.05). It provided significantly larger apparent signal-to-noise and apparent contrast-to-noise ratio values than gated CEMRA and unenhanced 3D bSSFP (P < 0.05) and provided ninefold better fluid suppression than equilibrium phase 3D bSSFP. Aortic diameter and main pulmonary artery diameter measurements obtained with bT1RESS and first-pass gated CEMRA strongly correlated (P < 0.05).

CONCLUSIONS:

We found that using bT1RESS greatly prolongs the useful duration of blood pool contrast enhancement while improving angiographic image quality compared with standard CEMRA techniques. Although further study is needed, potential advantages for vascular imaging include eliminating the current requirement for first-pass imaging along with better reliability and accuracy for a wide range of cardiovascular applications.