Accuracy of a novel real-time continuous glucose monitoring system: a prospective self-controlled study in thirty hospitalized patients with type 2 diabetes.
Ge, Shenghui; Zhang, Hui; Wang, Jun; Li, Huiqin; Su, Xiaofei; Ding, Dafa; Ma, Jianhua.
Afiliação
  • Ge S; Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
  • Zhang H; Department of Endocrine, Nanjing Tongren Hospital, School of Medicine, Southeast University, Nanjing, China.
  • Wang J; Department of Health Management Center, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
  • Li H; Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
  • Su X; Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
  • Ding D; Department of Endocrinology, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Ma J; Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing, China.
Front Endocrinol (Lausanne) ; 15: 1374496, 2024.
Article em En | MEDLINE | ID: mdl-38836229
ABSTRACT

Aims:

The present study aimed to investigate the accuracy of the Glunovo® real-time continuous glucose monitoring system (rtCGMS).

Methods:

We conducted a 14-day interstitial glucose level monitoring using Glunovo® rtCGMS on thirty hospitalized patients with type 2 diabetes. The flash glucose monitoring (FGM) was used as a self-control. Consistency tests, error grid analysis, and calculation of the mean absolute relative difference (MARD) were performed using R software to assess the accuracy of Glunovo® rtCGMS.

Results:

Glunovo® exhibited an overall MARD value of 8.89% during hospitalization, compared to 10.42% for FGM. The overall percentages of glucose values within ±10%/10, ± 15%/15, ± 20%/20, ± 30%/30, and ±40%/40 of the venous blood glucose reference value were 63.34%, 81.31%, 90.50%, 97.29%, and 99.36% for Glunovo®, respectively, compared with 61.58%, 79.63%, 88.31%, 96.22% and 99.23% for FGM. The Clarke Error Grid Analysis showed that 99.61% of Glunovo® glucose pairs and 100.00% of FGM glucose pairs within zones A and B.

Conclusion:

Our study confirms the superior accuracy of Glunovo® in monitoring blood glucose levels among hospitalized patients with type 2 diabetes.