Your browser doesn't support javascript.
loading
Early structural valve deterioration of balloon expandable transcatheter aortic valve leaflets due to intrinsic and extrinsic nodular calcifications in a haemodialysis patient: a case report.
Onishi, Kyohei; Mizutani, Kazuki; Sato, Yu; Nakazawa, Gaku.
Afiliação
  • Onishi K; Division of Cardiology, Department of Medicine, Kindai University Faculty of Medicine, 377-2 Ohno-Higashi, Osakasayama, Osaka 589-8511, Japan.
  • Mizutani K; Division of Cardiology, Department of Medicine, Kindai University Faculty of Medicine, 377-2 Ohno-Higashi, Osakasayama, Osaka 589-8511, Japan.
  • Sato Y; Department of Cardiology, Tokai University School of Medicine, Isehara, Japan.
  • Nakazawa G; Division of Cardiology, Department of Medicine, Kindai University Faculty of Medicine, 377-2 Ohno-Higashi, Osakasayama, Osaka 589-8511, Japan.
Eur Heart J Case Rep ; 8(6): ytae265, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38845807
ABSTRACT

Background:

Several reports have shown that transcatheter aortic valves are comparable in durability to surgical aortic valves. However, early structural valve deterioration (SVD) is rarely reported to occur, especially in haemodialysis patients. Case

summary:

We present a case of rapidly progressive bioprosthetic aortic valve stenosis in a patient with end-stage renal disease secondary to diabetic nephropathy in an 83-year-old female admitted due to progressive dyspnoea and orthopnoea. A 23 mm sized SAPIEN3 bioprosthetic aortic valve showed normal function for the first year after transcatheter aortic valve implantation (TAVI), but then rapidly developed stenosis and required acute hospitalization for heart failure a year and a half after surgery. Emergent surgical aortic valve replacement with a 19 mm On-X valve (CryoLife, Kennesaw, GA, USA) was performed. Pathological examination of the explanted SAPIEN 3 valve demonstrated severely degenerated bioprosthetic pericardial leaflets with severe intrinsic and extrinsic nodular calcifications, which could limit the leaflet motion.

Discussion:

There is a lack of reports on the long-term procedural outcomes of TAVI in haemodialysis patients. The development of SVD in patients undergoing dialysis is multifactorial and has yet to be fully elucidated. In the presented case, the removed TAVI valve had severe extrinsic calcified nodules alongside a fibrin thrombus. Considering these pathological findings, antithrombotic therapy to prevent fibrin thrombus from adhering to the TAVI valve may be important to avoid early SVD.