Your browser doesn't support javascript.
loading
Variations in emergency hemorrhage panel turnaround times in 2 major medical centers using the same laboratory methods.
Hogan, Matthew E; Liu, Zhinan; Stansbury, Lynn G; Vavilala, Monica S; Hess, John R; Tsang, Hamilton C.
Afiliação
  • Hogan ME; Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, US.
  • Liu Z; Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, US.
  • Stansbury LG; Harborview Injury Prevention Research Center, Harborview Medical Center, Seattle, WA, US.
  • Vavilala MS; Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, US.
  • Hess JR; Harborview Injury Prevention Research Center, Harborview Medical Center, Seattle, WA, US.
  • Tsang HC; Department of Anesthesia and Pain Medicine, University of Washington, Seattle, WA, US.
Am J Clin Pathol ; 2024 Jun 17.
Article em En | MEDLINE | ID: mdl-38884115
ABSTRACT

OBJECTIVES:

Demand for rapid coagulation testing for massive transfusion events led to development of an emergency hemorrhage panel (EHP; hemoglobin, platelet count, prothrombin time/international normalized ratio, and fibrinogen), with laboratory turnaround time (TAT) of less than 20 minutes. Ten years on, we asked if current laboratory practices were meeting that TAT goal and differences were evident in TAT between the 2 major institutions in our system.

METHODS:

We identified EHPs ordered at our 2 largest hospitals, February 2, 2021, to July 17, 2022, comparing order to specimen draw time, specimen draw to specimen received time, laboratory analytic time, and total TAT results from emergency department and operating room. Site 1 houses a level I trauma center; site 2 includes tertiary care, transplant, and obstetrics services.

RESULTS:

In total, 1137 EHPs were recorded in our study period. Laboratory TAT was significantly faster at site 1 (~14 vs ~27 minutes, P < .01). Average laboratory TAT was under 20 minutes at site 1 but only for 50% of specimens at site 2. Outlier specimens were collection delays at site 1 and specimen processing delays at site 2.

CONCLUSIONS:

The EHP can be performed as rapidly as described. However, compromises in laboratory location, available personnel, and processing differences can degrade performance.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article