Your browser doesn't support javascript.
loading
Development and validation of an automated Tomotherapy planning method for cervical cancer.
Han, Feiru; Xue, Yi; Huang, Sheng; Lu, Tong; Yang, Yining; Cao, Yuanjie; Chen, Jie; Hou, Hailing; Sun, Yao; Wang, Wei; Yuan, Zhiyong; Tao, Zhen; Jiang, Shengpeng.
Afiliação
  • Han F; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Xue Y; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Huang S; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Lu T; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Yang Y; Department of Radiation Oncology, Tianjin First Central Hospital, Tianjin, China.
  • Cao Y; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Chen J; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Hou H; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Sun Y; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Wang W; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Yuan Z; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China.
  • Tao Z; Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital; National Clinical Research Center for Cancer; Tianjin's Clinical Research Center for Cancer; Key Laboratory of Cancer Prevention and Therapy, Tianjin, China. ztao@tmu.edu.cn.
  • Jiang S; Department of Oncology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China. ztao@tmu.edu.cn.
Radiat Oncol ; 19(1): 88, 2024 Jul 08.
Article em En | MEDLINE | ID: mdl-38978062
ABSTRACT

PURPOSE:

This study aimed to develop an automated Tomotherapy (TOMO) planning method for cervical cancer treatment, and to validate its feasibility and effectiveness. MATERIALS AND

METHODS:

The study enrolled 30 cervical cancer patients treated with TOMO at our center. Utilizing scripting and Python environment within the RayStation (RaySearch Labs, Sweden) treatment planning system (TPS), we developed automated planning methods for TOMO and volumetric modulated arc therapy (VMAT) techniques. The clinical manual TOMO (M-TOMO) plans for the 30 patients were re-optimized using automated planning scripts for both TOMO and VMAT, creating automated TOMO (A-TOMO) and automated VMAT (A-VMAT) plans. We compared A-TOMO with M-TOMO and A-VMAT plans. The primary evaluated relevant dosimetric parameters and treatment plan efficiency were assessed using the two-sided Wilcoxon signed-rank test for statistical analysis, with a P-value < 0.05 indicating statistical significance.

RESULTS:

A-TOMO plans maintained similar target dose uniformity compared to M-TOMO plans, with improvements in target conformity and faster dose drop-off outside the target, and demonstrated significant statistical differences (P+ < 0.01). A-TOMO plans also significantly outperformed M-TOMO plans in reducing V50Gy, V40Gy and Dmean for the bladder and rectum, as well as Dmean for the bowel bag, femoral heads, and kidneys (all P+ < 0.05). Additionally, A-TOMO plans demonstrated better consistency in plan quality. Furthermore, the quality of A-TOMO plans was comparable to or superior than A-VMAT plans. In terms of efficiency, A-TOMO significantly reduced the time required for treatment planning to approximately 20 min.

CONCLUSION:

We have successfully developed an A-TOMO planning method for cervical cancer. Compared to M-TOMO plans, A-TOMO plans improved target conformity and reduced radiation dose to OARs. Additionally, the quality of A-TOMO plans was on par with or surpasses that of A-VMAT plans. The A-TOMO planning method significantly improved the efficiency of treatment planning.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Neoplasias do Colo do Útero / Radioterapia de Intensidade Modulada / Órgãos em Risco Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Planejamento da Radioterapia Assistida por Computador / Neoplasias do Colo do Útero / Radioterapia de Intensidade Modulada / Órgãos em Risco Idioma: En Ano de publicação: 2024 Tipo de documento: Article