Utility of block-matching and 3D filter for reproducibility of lung density and denoising in low-dose chest CT: A pilot study.
Shim, Jina; Kang, Seong-Hyeon; Lee, Youngjin.
Afiliação
  • Shim J; Department of Diagnostic Radiology, Severance Hospital, Seoul, Republic of Korea.
  • Kang SH; Department of Radiological Science, Gachon University, Incheon, Republic of Korea. Electronic address: tjdgus7345@gachon.ac.kr.
  • Lee Y; Department of Radiological Science, Gachon University, Incheon, Republic of Korea. Electronic address: yj20@gachon.ac.kr.
Phys Med ; 124: 103432, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38996628
ABSTRACT

PURPOSE:

This study aimed to acquire an image quality consistent with that of full-dose chest computed tomography (CT) when obtaining low-dose chest CT images and to analyze the effects of block-matching and 3D (BM3D) filters on lung density measurements and noise reduction in lung parenchyma.

METHODS:

Using full-dose chest CT images, we evaluated lung density measurements and noise reduction in lung parenchyma images for low-dose chest CT. Three filters (median, Wiener, and the proposed BM3D) were applied to low-dose chest CT images for comparison and analysis with images from full-dose chest CT. To evaluate lung density measurements, we measured CT attenuation at the 15th percentile of the lung CT histogram. The coefficient of variation (COV) and contrast-to-noise ratio (CNR) were used to evaluate the noise level.

RESULTS:

The 15th percentile of the lung CT histogram showed the smallest difference between full- and low-dose CT when applying the BM3D filter, and the highest difference between full- and low-dose CT without filters (full-dose =  - 926.28 ± 0.32, BM3D =  - 926.65 ± 0.32, and low-dose =  - 959.43 ± 0.95) (p < 0.05). The COV was smallest when applying the BM3D filter, whereas the CNR was the highest (p < 0.05).

CONCLUSIONS:

The results of the study prove that the BM3D filter can reduce image noise while increasing the reproducibility of the lung density, even for low-dose chest CT.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Doses de Radiação / Tomografia Computadorizada por Raios X / Razão Sinal-Ruído / Pulmão Idioma: En Ano de publicação: 2024 Tipo de documento: Article