90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit.
Lopez, Benjamin P; Kappadath, S Cheenu.
Afiliação
  • Lopez BP; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas.
  • Kappadath SC; Department of Imaging Physics, University of Texas MD Anderson Cancer Center, Houston, Texas skappadath@mdanderson.org.
J Nucl Med Technol ; 2024 Jul 17.
Article em En | MEDLINE | ID: mdl-39019573
ABSTRACT
A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit.

Methods:

Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified.

Results:

SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (R 2 = 0.985).

Conclusion:

Dose calibrator settings and exposure-rate-to-activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.