Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation dose spine computed tomography (CT) images.
Sun, Jihang; Li, Haoyan; Yu, Tong; Huo, Aihua; Hua, Shan; Zhou, Zuofu; Peng, Yun.
Afiliação
  • Sun J; Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
  • Li H; Medical Imaging Department, Children's Hospital of Xinjiang Uygur Autonomous Region, Xinjiang Hospital of Beijing Children's Hospital, Urumqi, China.
  • Yu T; Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
  • Huo A; Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
  • Hua S; Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
  • Zhou Z; Medical Imaging Department, Children's Hospital of Xinjiang Uygur Autonomous Region, Xinjiang Hospital of Beijing Children's Hospital, Urumqi, China.
  • Peng Y; Department of Radiology, Fujian Maternity and Child Health Hospital College of Clinical Medicine for Obstetrics & Gynecology and Pediatrics, Fujian Medical University, Fujian, China.
Quant Imaging Med Surg ; 14(7): 4648-4658, 2024 Jul 01.
Article em En | MEDLINE | ID: mdl-39022236
ABSTRACT

Background:

The commonly used methods for removing metal-induced beam hardening artifacts often rely on the use of high energy photons with either high tube voltage or high energy virtual monoenergetic images in dual-energy computed tomography (CT), the radiation dose was usually relatively high in order to generate adequate signals. This retrospective study is designed to evaluate the application of a metal artifact reduction (MAR) algorithm in reducing pedicle screw metal-caused beam hardening artifacts in post-surgery pediatric low radiation dose spine CT images.

Methods:

Seventy-seven children (3-15 years) who had undergone a low dose CT with 140 or 100 kV were enrolled. The radiation dose was 1.40 mGy for the 3-8 years old and 2.61 mGy for 9-15 years old children. There were 116 pedicle screws evaluated. The raw data were reconstructed with adaptive statistical iterative reconstruction-V (ASIR-V) at 50% strength, ASIR-V with MAR (AV-MAR), deep learning image reconstruction (DLIR) at high strength and DLIR with MAR (DL-MAR). The image quality concerning pedicle screws was evaluated objectively in terms of the length of beam-hardening artifact (LHA) and artifact index (AI), and subjectively using a 4-point scale (4 points best, 3 points acceptable).

Results:

Both AV-MAR and DL-MAR significantly reduced metal-induced beam hardening artifacts with smaller LHA (15.76±10.12 mm, a reduction of 57.24% and 15.66±10.49 mm, a reduction of 57.40%, respectively), and AI value (62.50±33.51, a reduction of 64.65% and 61.03±32.61, a reduction of 65.01%, respectively) compared to ASIR-V and DLIR (all P<0.01), The subjective image quality scores concerning the screws were 3.37±0.49 and 3.47±0.50 with AV-MAR and DL-MAR, respectively, higher than the respective value of 1.73±0.44 and 1.76±0.43 without MAR (all P<0.01).

Conclusions:

MAR significantly reduces the low-density artifacts caused by metal screws in post-surgery pediatric low-dose spine CT images, across different tube voltages, radiation dose levels and reconstruction algorithms. Combining DL-MAR further improves the overall image quality under low radiation dose conditions.