Your browser doesn't support javascript.
loading
Ultrasound propagation characteristics within the bone tissue of miniature ultrasound probes: implications for the spinal navigation of pedicle screw placement.
Li, Peiyang; Li, Suoyuan; Li, Zhiqiang; Lu, Wenjia; Shao, Weiwei; Li, Zhangjian; Xu, Yiwen; Zhang, Hong; Ju, Bin; Shen, Jun; Cui, Yaoyao.
Afiliação
  • Li P; Academy for Engineering & Technology, Fudan University, Shanghai, China.
  • Li S; Department of Orthopeadics, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Gusu School, Nanjing Medical University, Suzhou, China.
  • Li Z; Department of Orthopeadics, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Gusu School, Nanjing Medical University, Suzhou, China.
  • Lu W; Suzhou GuoKe Ultra Medical Technology Co., Ltd., Suzhou, China.
  • Shao W; Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou, China.
  • Li Z; Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou, China.
  • Xu Y; Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou, China.
  • Zhang H; Department of Orthopeadics, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Gusu School, Nanjing Medical University, Suzhou, China.
  • Ju B; College of Electrical Engineering and Automation, Anhui University, Hefei, China.
  • Shen J; Department of Orthopeadics, The Affiliated Suzhou Hospital of Nanjing Medical University, Suzhou Municipal Hospital, Gusu School, Nanjing Medical University, Suzhou, China.
  • Cui Y; Suzhou Key Laboratory of Orthopedic Medical Engineering, Suzhou, China.
Quant Imaging Med Surg ; 14(7): 4878-4892, 2024 Jul 01.
Article em En | MEDLINE | ID: mdl-39022289
ABSTRACT

Background:

The accuracy of pedicle screw fixation is crucial for patient safety. Traditional navigation methods based on computed tomography (CT) imaging have several limitations. Therefore, this study aimed to investigate the ultrasonic propagation characteristics of bone tissue and their relationship with CT imaging results, as well as the potential application of ultrasound navigation in pedicle screw fixation.

Methods:

The study used three bovine spine specimens (BSSs) and five human vertebral allograft bones (HABs) to progressively decrease the thickness of the cancellous bone layer, simulating the process of pedicle screw perforation. Five unfocused miniature ultrasound probes with frequencies of 2.2, 2.5, 3, 12, and 30 MHz were employed for investigating the ultrasonic propagation characteristics of cancellous and cortical bone through ultrasound transmission and backscatter experiments. The CT features of the bone tissue was obtained with the Skyscan 1174 micro-CT scanner (Bruker, Billerica, MA, USA).

Results:

The experimental results demonstrated that low-frequency (2-3 MHz) ultrasound effectively penetrated the cancellous bone layer up to a depth of approximately 5 mm, with an attenuation coefficient below 10 dB/cm. Conversely, high-frequency (12 MHz) ultrasound exhibited significant signal attenuation in cancellous bone, reaching up to 55.8 dB/cm. The amplitude of the backscattered signal at the cancellous bone interface exhibited a negative correlation with the bone sample thickness (average r=-0.84), meaning that as the thickness of the cancellous bone layer on the cortical bone decreases, the backscattered signal amplitude gradually increases (P<0.05). Upon reaching the cortical bone interface, there was a rapid surge in echo signal amplitude, up to 8 times higher. Meanwhile, the statistical results indicated a significant correlation between the amplitude of the echo signal and the micro-CT scanning results of bone trabecular structure.

Conclusions:

Theoretically, using multiple ultrasonic probes (≥3) and regions of interest (ROIs) (≥5) has the potential to provide surgeons with early warning signals for pedicle perforation based on three or more successive increases in echo signal amplitude or a sudden substantial increase. The statistical results indicate a significant correlation between the amplitude of the echo signal and the micro-CT scanning results of bone trabeculae, suggesting the potential use of ultrasound as opposed to CT for real-time intraoperative bone navigation.
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article