Tremor Estimation and Removal in Robot-Assisted Surgery Using Improved Enhanced Band-Limited Multiple Fourier Linear Combiner.
Int J Med Robot
; 20(4): e2666, 2024 Aug.
Article
em En
| MEDLINE
| ID: mdl-39092625
ABSTRACT
BACKGROUND:
During a robot-assisted minimally invasive surgery, hand tremors in a surgeon's manipulation of the master manipulator can cause vibrations of the slave surgical instruments.METHODS:
This letter addresses this problem by proposing an improved Enhanced Band-Limited Multiple Linear Fourier Combiner (E-BMFLC) algorithm for filtering the physiological tremor signals of a surgeon's hand. The proposed method uses the amplitude of the input signal to adapt the learning rate and a dense division of the combiner bands for the higher amplitude bands of the tremor signals.RESULTS:
By using the proposed improved E-BMFLC algorithm, the compensation accuracy can be improved by 4.5%-8.9%, as well as a spatial position error of less than 1 mm.CONCLUSION:
The results show that among all filtering methods, the improved E-BMFLC filtering method has the highest number of successful experiments and the lowest experimental time.Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Tremor
/
Algoritmos
/
Procedimentos Cirúrgicos Robóticos
/
Análise de Fourier
Idioma:
En
Ano de publicação:
2024
Tipo de documento: