Determination of chlordimeform and its metabolite residue in milk by gas chromatography-tandem mass spectrometry.
Food Res Int
; 192: 114754, 2024 Sep.
Article
em En
| MEDLINE
| ID: mdl-39147558
ABSTRACT
Herein, the determination of chlordimeform and its metabolite 4-chloro-2-methylaniline residue in milk was performed for the first time using gas chromatography-tandem mass spectrometry (GC-MS/MS). Samples were extracted using acetonitrile, and cleaned using a quick, easy, cheap, effective, rugged, and safe (QuEChERS) method. Separation was performed using the DB-17 MS column. It was detected in a selected reaction monitoring (SRM) mode and quantified using a matrix-matched isotope internal standard method. Under optimal conditions, a good linear relationship was observed in the concentration range of 10-200 µg/kg. The limit of quantitation was 10.0 µg/kg. The spiked recoveries for the target substance ranged from 84.5 % to 107.3 %, with relative standard deviations (RSD) of <7.2 %. The spiked samples were further confirmed by gas chromatography-quadrupole-Orbitrap high-resolution mass spectrometry (GC-Orbitrap HRMS). The combined method resulted in high accuracy and sensitivity and was suitable for the determination of chlordimeform and its metabolite 4-chloro-2-methylaniline residue in milk.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Contaminação de Alimentos
/
Leite
/
Espectrometria de Massas em Tandem
/
Cromatografia Gasosa-Espectrometria de Massas
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article