The multi-residue determination of coumarin-based anticoagulant rodenticides in animal materials by high-performance liquid chromatography.
J Chromatogr ; 234(2): 427-35, 1982 Jan 15.
Article em En | MEDLINE | ID: mdl-7056833
ABSTRACT
The rodenticides brodifacoum, difenacoum, coumatetralyl and warfarin are determined in animal relicta by high-performance exclusion chromatography on porous silica. The first three compounds are not separated, but are subsequently differentiated by adsorption or reversed-phase high-performance liquid chromatography of the appropriate eluate fraction collected from the exclusion column. The method is rapid, and clean-up (on Sep-Pak silica cartridges) is simple. Mean recoveries from spiked substrates were generally above 80% at levels of 0.1-1.0 mg/kg. Routine limits of determination are about 0.05-0.1 mg/kg for warfarin and about 0.02 mg/kg for the other compounds. If analysis for warfarin is not required, the latter limit can be lowered to about 1 microgram/kg by a slight modification to the clean-up step.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Rodenticidas / 4-Hidroxicumarinas Idioma: En Ano de publicação: 1982 Tipo de documento: Article
Buscar no Google