Single X-ray absorptiometry of the forearm: precision, correlation, and reference data.
Kelly, T L; Crane, G; Baran, D T.
Afiliação
  • Kelly TL; Hologic, Inc., Waltham, Massachusetts 02154.
Calcif Tissue Int ; 54(3): 212-8, 1994 Mar.
Article em En | MEDLINE | ID: mdl-8055369
ABSTRACT
The performance of a single X-ray absorptiometry (SXA) device incorporating an X-ray tube as a photon source was evaluated with respect to precision in vivo and in vitro, scan time, image quality, and correlation with an existing dual energy X-ray absorptiometry (DXA) device. SXA precision in vivo, expressed as a coefficient of variation (CV), was 0.66% for bone mineral content (BMC) and 1.05% for bone mineral density (BMD). Precision in vitro, based on 78 BMC measurements of a forearm phantom over 195 days, was 0.53%. Correlation with DXA at the 8 mm distal forearm site was high (r = 0.97 for BMC and r = 0.96 for BMD). A preliminary SXA reference database composed of 151 healthy Caucasian American women was developed to facilitate the interpretation of patient measurements. SXA scan time was 4 minutes and delivered a radiation exposure of 1.68 mrem. SXA image quality and spatial resolution were superior to SPA and comparable to DXA.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Absorciometria de Fóton / Densidade Óssea Idioma: En Ano de publicação: 1994 Tipo de documento: Article