An example of thermometry in volume by microwave radiometry.
Bocquet, B; Van de Velde, J C; Mamouni, A; LeRoy, Y.
Afiliação
  • Bocquet B; Institut d'Electronique et de Microélectronique du Nord, Centre Hyperfrequences et Semiconducteurs, Université des Sciences et Technologies de Lille, Villeneuve d'Ascq, France.
IEEE Trans Biomed Eng ; 40(9): 990-2, 1993 Sep.
Article em En | MEDLINE | ID: mdl-8288291
ABSTRACT
We depict a method of determination of the size D depth z and temperature T0 + delta T of a cylindrical thermal structure embedded in an homogeneous glossy material, in the present case, water. A microwave radiometric image at 3 GHz points out the location of the thermal structure; its threshold provides the diameter D of the structure. The depth z derives from the ratio of the maximal radiometric intensities at 1.5 and 3 GHz. The combination of D,z and of the radiometric intensities gives delta T.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Termômetros / Processamento de Imagem Assistida por Computador / Micro-Ondas Idioma: En Ano de publicação: 1993 Tipo de documento: Article