Your browser doesn't support javascript.
loading
[Ultra-low-dose spiral (helical) CT of the thorax: a filtering technique].
Nitta, N; Takahashi, M; Murata, K; Mori, M; Shimoyama, K; Mishina, A; Matsuo, H; Morita, R; Sugii, K; Nomura, A.
Afiliação
  • Nitta N; Department of Radiology, Shiga University of Medical Science.
Nihon Igaku Hoshasen Gakkai Zasshi ; 56(1): 63-5, 1996 Jan.
Article em Ja | MEDLINE | ID: mdl-8857102
To reduce the radiation dose from spiral (helical) CT, a custom-made aluminium filter was installed in the X-ray tube and a reduction of effective tube current was attempted. A pronounced reduction of effective tube current, namely, 6 and 3 mA, was achieved with 26 and 37 mm thick aluminium filters, respectively. Visualization of normal lung structure was accomplished with both 6 and 3 mA settings. However, images of 3 mA failed to delineate mediastinal structures because of marked beam hardening resulting from the bone structure of the thoracic inlet. Six mA was considered the lowest dose setting of spiral (helical) CT of the thorax that could be used for lung cancer screening.
Assuntos
Buscar no Google
Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Tomografia Computadorizada por Raios X / Pulmão Idioma: Ja Ano de publicação: 1996 Tipo de documento: Article
Buscar no Google
Base de dados: MEDLINE Assunto principal: Proteção Radiológica / Tomografia Computadorizada por Raios X / Pulmão Idioma: Ja Ano de publicação: 1996 Tipo de documento: Article