1.
Rev Sci Instrum ; 94(7)2023 Jul 01.
Artículo en Inglés | MEDLINE | ID: mdl-37493502

RESUMEN

We describe the results of our investigations of the generation of a ribbon electron beam (10 × 220 mm2) by a two-stage discharge system based on a hollow-cathode glow discharge plasma. The source design enables operation in the pressure range 2 × 10-2 to 10 Pa. At a beam accelerating voltage of 8 kV, the beam current is 450 mA at a pressure of 2 × 10-2 Pa and 150 mA at a pressure of 10 Pa. To achieve a uniform current density distribution of the beam over its cross-sectional area, a special design of emission electrode was employed. This enabled us to reduce non-uniformities of the beam current density distribution to a level of 10%.

2.
Rev Sci Instrum ; 85(2): 02A738, 2014 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-24593472

RESUMEN

The paper describes the principle of operation, design special features, and parameters of an inverted time-of-flight spectrometer. The spectrometer is designed in such way that its deflecting plates, drift tube, and primary measuring system are at high potential with respect to the ground potential, whereas plasma is formed near grounded electrodes. This type of configuration greatly extends the application range of the device, making it possible to measure the mass-to-charge composition of plasma with wide range of parameters.