Transfusion ; 57(1): 122-130, 2017 01.
Artículo en Inglés | MEDLINE | ID: mdl-27696435

RESUMEN

BACKGROUND: Peripheral stem cell collections can be challenging in the pediatric population and respective experience is limited. Since February 2015 our institution is utilizing the new Spectra Optia (Optia) apheresis device, which has replaced the former COBE Spectra (COBE) device. As a quality initiative we collected and compared collection efficiency (CE2) and other collection variables between the two devices. STUDY DESIGN AND METHODS: In this retrospective study we collected and compared clinical, laboratory, and technical collection data from stem cell collection procedures done with the Optia and COBE devices. The collected data included patient demographics, precollection peripheral CD34+ cell counts, total CD34+ cells collected, complete blood count, electrolytes before and after collection, side effects attributed to the collection, total blood volumes processed (TBVs), collection times, and calculated CE2 and collection ratios. RESULTS: Forty-one collection procedures performed on 29 pediatric patients with the Optia device were compared to 41 collections performed on 27 patients with the COBE device. The TBVs through the Optia device were significantly smaller than the COBE (3.9 ± 0.2 × TBV vs. 5.5 ± 0.1 × TBV, respectively; p < 0.001), requiring significantly less anticoagulant and providing similar amounts of stem cells while collection times were significantly shorter (mean, 238 ± 9 min vs. 264 ± 9 min, respectively; p < 0.05). Collections on the Optia caused significantly smaller reductions of plasma calcium and magnesium. No significant side effects attributed to the procedure were noted. CONCLUSION: Stem cell apheresis with the Optia device in children is safe and feasible with smaller blood volumes with shorter collection times.


Asunto(s)
Eliminación de Componentes Sanguíneos/instrumentación , Eliminación de Componentes Sanguíneos/métodos , Neoplasias/terapia , Trasplante de Células Madre de Sangre Periférica , Células Madre de Sangre Periférica , Adolescente , Autoinjertos , Niño , Preescolar , Femenino , Humanos , Lactante , Masculino , Neoplasias/sangre , Estudios Retrospectivos , Factores de Tiempo
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA