J Laryngol Otol ; 132(10): 943-945, 2018 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-30152297

RESUMEN

BACKGROUND: Dislocation of the cochlear implant magnet is an uncommon but recognised complication of magnetic resonance imaging in a patient with a cochlear implant.Case reportThis paper reports a case of cochlear implant magnet dislocation despite head bandaging. The patient subsequently underwent endoscopic repositioning of the magnet under general anaesthesia. The cochlear implant system was tested intra-operatively and confirmed to be functioning. The patient was well at follow up and the small wound healed well with no complications. CONCLUSION: This report presents the endoscopic technique as a viable minimally invasive surgical approach to address cochlear implant magnet dislocation.


Asunto(s)
Implantación Coclear , Implantes Cocleares , Falla de Equipo , Imagen por Resonancia Magnética , Cirugía Endoscópica por Orificios Naturales/métodos , Posicionamiento del Paciente , Humanos , Imanes , Reoperación , Resultado del Tratamiento