RESUMEN
AIMS: Wearable cardioverter-defibrillators (WCDs) are indicated in patients at risk of sudden cardiac arrest who are not immediate candidates for implantable defibrillator therapy. Limitations of existing WCDs include poor compliance and high false alarm rates. The Jewel is a novel patch-WCD (P-WCD) that addresses these limitations with an adhesive-based design for near-continuous wear and a machine learning algorithm designed to minimize inappropriate detections. This was a first-in-human study of the Jewel P-WCD conducted in an electrophysiology (EP) lab to determine the safety and effectiveness of the device in terminating ventricular tachycardia/ventricular fibrillation (VT/VF) with a single shock. The aim was to evaluate the safety and effectiveness of terminating VT/VF with a single shock using the Jewel P-WCD. METHODS AND RESULTS: This was a first-in-human, prospective, single-arm, single-centre study in patients scheduled for an EP procedure in which VT/VF was expected to either spontaneously occur or be induced. The Jewel P-WCD was placed on consented patients; upon confirmation of VT/VF, a single shock (150 J) was delivered via the device. A group sequential design and Pocock alpha spending function was used to measure the observed proportion of successful VT/VF single-shock terminations. The endpoint was achieved if the lower confidence limit exceeded the performance goal of 62%, using a one-sided lower 97.4% exact confidence bound. Of 18 eligible subjects, 16 (88.9%, 97.4% confidence bound: 65.4%) were successfully defibrillated with a single shock, exceeding the primary endpoint performance goal with no adverse events. CONCLUSION: This first-in-human evaluation of the Jewel P-WCD demonstrated the safety and effectiveness of terminating VT/VF. CLINICAL TRIAL REGISTRATION: URL: https://clinicaltrials.gov/; Unique identifier: NCT05490459.
Asunto(s)
Desfibriladores , Cardioversión Eléctrica , Taquicardia Ventricular , Fibrilación Ventricular , Dispositivos Electrónicos Vestibles , Humanos , Masculino , Femenino , Fibrilación Ventricular/terapia , Fibrilación Ventricular/diagnóstico , Cardioversión Eléctrica/instrumentación , Cardioversión Eléctrica/efectos adversos , Persona de Mediana Edad , Taquicardia Ventricular/terapia , Taquicardia Ventricular/diagnóstico , Taquicardia Ventricular/fisiopatología , Estudios Prospectivos , Resultado del Tratamiento , Anciano , Diseño de Equipo , Adulto , Muerte Súbita Cardíaca/prevención & controlRESUMEN
OBJECTIVES: To investigate intrafascicular pudendal nerve stimulation in felines as a means to restore urinary function in acute models of urinary incontinence, overactive bladder, and underactive bladder. MATERIALS AND METHODS: Felines were anesthetized, and high-electrode-count (48 electrodes; 25 electrodes/mm(2) ) electrode arrays were implanted intrafascicularly into the pudendal nerve trunk. Electrodes were mapped for their ability to selectively or nonselectively excite the external anal sphincter, external urethral sphincter, and the detrusor bladder muscle. Statistical analysis was carried out to quantify reflexive voiding efficiencies, mean impedances of the microelectrodes used in this study, and to determine what differences, if any, in bladder contraction amplitudes were evoked by different electrode configurations. RESULTS: Multielectrode arrays implanted into the pudendal nerve trunk were able to selectively and nonselectively excite genitourinary muscles. After inducing urinary incontinence with bilateral pudendal nerve transections (proximal to the implants), electrical stimulation delivered through certain microelectrodes was able to significantly reduce leaking (p = 0.008). Electrical stimulation delivered through detrusor selective electrodes was able to inhibit reflexive bladder contractions and excite bladder contractions, depending on the stimulation frequency. Specific electrode configurations were able to drive significantly (p < 0.001) larger bladder contractions than other electrode configurations, depending on the preparation. Successful reflexively or electrically driven bladder contractions were achieved in 46% and 38% of the preparations, respectively, an observation that has not been noted in previously published feline pudendal stimulation studies. CONCLUSIONS: Multielectrode arrays implanted intrafascicularly into the pudendal nerve trunk may provide a promising new clinical neuromodulation therapy for the restoration of urinary function.
Asunto(s)
Terapia por Estimulación Eléctrica/métodos , Electrodos Implantados , Nervio Pudendo/fisiología , Recuperación de la Función/fisiología , Trastornos Urinarios/terapia , Animales , Gatos , Modelos Animales de Enfermedad , Masculino , Contracción Muscular/fisiología , Músculo Liso/fisiología , Reflejo/fisiología , Vejiga Urinaria Hiperactiva/terapia , Incontinencia Urinaria/terapiaRESUMEN
INTRODUCTION: High density Utah slanted electrode arrays (HD-USEAs) have been developed recently for intrafascicular access to submillimeter neural structures. Insertion of such high electrode density devices may cause nerve crush injury, counteracting the intended improved selective nerve fiber access. METHODS: HD-USEAs were implanted into sciatic nerves of anesthetized rats. Nerve function was assessed before and after HD-USEA implantation by measuring changes in evoked muscle and nerve compound action potentials and single unit neuronal recordings. RESULTS: Neural activity was recorded with over half of all implanted electrodes. Average decreases of 38%, 36%, and 13% in nerve, medial gastrocnemius, and tibialis anterior compound action potential amplitudes, respectively, were observed following array implantation. Only 1 of 8 implantations resulted in loss of all signals. CONCLUSIONS: These studies demonstrate that HD-USEAs provide a useful neural interface without causing a nerve crush injury that would otherwise negate their use in acute preparations (<12 h).
Asunto(s)
Electrodos Implantados , Músculo Esquelético/fisiología , Nervio Ciático/fisiología , Potenciales de Acción/fisiología , Animales , Fenómenos Electrofisiológicos/fisiología , Potenciales Evocados/fisiología , Pie/inervación , Pie/fisiología , Microelectrodos , Ratas , Ratas Sprague-DawleyRESUMEN
BACKGROUND: For many patients, sudden cardiac arrest (SCA) risk is elevated temporarily. Wearable cardioverter-defibrillators (WCDs) can monitor and treat SCA during these temporary periods. Traditional WCDs can be uncomfortable, require frequent maintenance, and cannot be used when showering, resulting in poor compliance and avoidable SCA deaths. The Jewel is a novel, water-resistant patch-wearable cardioverter-defibrillator (P-WCD) with a machine learning detection algorithm designed to improve compliance and protection against SCA. OBJECTIVES: This study aims to demonstrate the safety and clinical effectiveness of a novel P-WCD. METHODS: The Jewel IDE Study, a prospective, single-arm study conducted at 30 U.S. sites, enrolled patients at SCA risk due to ventricular tachycardia/ventricular fibrillation who were not candidates for or refused an implantable defibrillator. The primary safety endpoint was <15% patients with clinically significant cutaneous adverse device effects and the primary effectiveness endpoint was <2 inappropriate shocks/100 patient-months. Secondary endpoints were ≥1 successful ventricular tachycardia/ventricular fibrillation conversion and wear time compliance of >14.1 h/d. RESULTS: A total of 305 patients (mean age: 57.9 years; 30.2% female, 27.9% non-White) were enrolled, of which 290 had available device data. The clinically significant cutaneous adverse device effect rate was 2.30% (upper 1-sided 98% CI: 4.80); none were severe. No device-related deaths or serious adverse events were reported. The inappropriate shock rate was 0.36/100 patient-months (upper 1-sided 98% CI: 1.53). Of 11 shocks in 9 patients, 9 shocks were adjudicated to be appropriate. Eight of 9 shocks were successful with a single shock. Median wear time compliance was 23.5 (20.7-23.9) h/d. CONCLUSIONS: The novel P-WCD is a safe and effective WCD with high patient compliance. There were no deaths due to noncompliance and a high number of successful conversions (Jewel IDE study [A Clinical Evaluation of the Jewel P-WCD in Subjects at High Risk for Sudden Cardiac Arrest]; NCT05201495).
Asunto(s)
Muerte Súbita Cardíaca , Desfibriladores , Dispositivos Electrónicos Vestibles , Humanos , Muerte Súbita Cardíaca/prevención & control , Muerte Súbita Cardíaca/epidemiología , Muerte Súbita Cardíaca/etiología , Femenino , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Anciano , Taquicardia Ventricular/terapia , Fibrilación Ventricular/terapia , Diseño de Equipo , Adulto , Cooperación del PacienteRESUMEN
INTRODUCTION: Injury mechanics of blunt impact projectiles differ from those experienced in whole body motor vehicle collisions because the effects are localized around the point of impact, and thus, injury thresholds based upon gross chest kinematics (e.g., force, velocity) may not be applicable across impact types. Therefore, knowledge of biomechanically based tissue injury correlates for blunt impact projectiles are needed to better guide design and development of protective systems as well as assess injury risks from blunt impact projectile weapons. MATERIALS AND METHODS: In this study, subject-specific swine finite element models were used to quantify the tissue-level stresses and strains resulting from high speed projectile impact. These tissue-level injury doses were correlated to pathology injury outcomes to produce injury risk curves for lung contusion. Details of the pathology data and finite element results are provided in Appendix 1. Survival analysis regression methods were applied to develop lung injury regression curves and a number of statistical methods were used to evaluate several biomechanical metrics as correlates to lung contusion. Uncertainty and sensitivity analyses were used to further confirm the selection of the correlate. RESULTS: Statistical analysis revealed that normalized strain-energy density was the best correlate for prediction of lung tissue damage. Going further, normalized strain-energy density also proved to be suitable for prediction of the percentage of contused lung volume, a more meaningful medical diagnosis. As expected, peak strain-energy density is most sensitive to muscle-skin properties, as quantified through a comprehensive uncertainty and sensitivity analysis over three sets of projectile weights and speeds. CONCLUSIONS: Normalized strain-energy density was found to be the best correlate for prediction of lung tissue damage and correlate well to extent of contused lung volume.
Asunto(s)
Contusiones , Lesión Pulmonar , Animales , Fenómenos Biomecánicos , Pulmón , PorcinosRESUMEN
INTRODUCTION: Blast-induced mild traumatic brain injury was generated in a mouse model using a shock tube to investigate recovery and axonal injury from single blast. METHODS: A supersonic helium wave hit the head of anesthetized male young adult mice with a reflected pressure of 69 psi for 0.2 ms on Day 1. Subsequently, the mice were cardioperfused on Days 2, 5, or 12. The isolated brains were subjected to diffusion tensor imaging. Reduced fractional anisotropy (FA) indicated axonal injury. RESULTS: After single blast, FA showed a biphasic response in the corpus callosum with decrease on Days 2 and 12 and increase on Day 5. CONCLUSIONS: Blast-induced mild traumatic brain injury in a mouse model follows a biphasic FA response within 12 days after a single blast similar to that reported for human subjects.
Asunto(s)
Anisotropía , Traumatismos por Explosión/complicaciones , Conmoción Encefálica/etiología , Animales , Traumatismos por Explosión/fisiopatología , Conmoción Encefálica/fisiopatología , Imagen de Difusión Tensora/métodos , Modelos Animales de Enfermedad , Explosiones/estadística & datos numéricos , RatonesRESUMEN
OBJECTIVE: To investigate the use of a microelectrode array with a high spatial density of penetrating intrafascicular electrodes for selective recording of pudendal nerve activity evoked by a variety of genitourinary stimuli. MATERIALS AND METHODS: Felines were anesthetized with alpha-chloralose and high-density Utah slanted electrode arrays (48 microelectrodes; 200-µm spacing) were implanted into the pudendal nerve for acute experimentation. Neural activity was recorded during bladder filling, spontaneous reflexive distention-evoked bladder contractions, and tactile somatosensory stimulation. RESULTS: The intrafascicularly implanted pudendal nerve electrodes were able to selectively record neural activity that corresponded to various genitourinary stimuli. Across all seven experimental animals, a total of 10 microelectrodes recorded neural units that were selectively driven by bladder filling or distention-evoked bladder contractions. Twenty-two electrodes were selectively driven by tactile stimulation. CONCLUSION: Microelectrode arrays implanted intrafascicularly into the pudendal nerve can be used to selectively record the neural responses that reflect bladder status and urogenital tactile stimulation. This work sets the stage for developing future implantable closed-loop neuroprosthetic devices for restoration of bladder function.