Lab Med ; 52(5): e137-e146, 2021 Sep 01.
Artículo en Inglés | MEDLINE | ID: mdl-33929022

RESUMEN

OBJECTIVE: To describe a cross-institutional approach to verify the Abbott ARCHITECT SARS-CoV-2 antibody assay and to document the kinetics of the serological response. METHODS: We conducted analytical performance evaluation studies using the Abbott ARCHITECT SARS-CoV-2 antibody assay on 5 Abbott ARCHITECT i2000 automated analyzers at 2 academic medical centers. RESULTS: Within-run and between-run coefficients of variance (CVs) for the antibody assay did not exceed 5.6% and 8.6%, respectively, for each institution. Quantitative and qualitative results agreed for lithium heparin plasma, EDTA-plasma and serum specimen types. Results for all SARS-CoV-2 IgG-positive and -negative specimens were concordant among analyzers except for 1 specimen at 1 institution. Qualitative and quantitative agreement was observed for specimens exchanged between institutions. All patients had detectable antibodies by day 10 from symptom onset and maintained seropositivity throughout specimen procurement. CONCLUSIONS: The analytical performance characteristics of the Abbott ARCHITECT SARS-CoV-2 antibody assay within and between 2 academic medical center clinical laboratories were acceptable for widespread clinical-laboratory use.


Asunto(s)
Anticuerpos Antivirales/sangre , Prueba Serológica para COVID-19/normas , COVID-19/diagnóstico , Inmunoensayo/normas , Inmunoglobulina G/sangre , SARS-CoV-2/inmunología , Centros Médicos Académicos , COVID-19/sangre , COVID-19/inmunología , COVID-19/virología , Humanos , Variaciones Dependientes del Observador , Reproducibilidad de los Resultados , SARS-CoV-2/patogenicidad , Sensibilidad y Especificidad , Virginia
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA