Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
País/Região como assunto
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
Radiat Prot Dosimetry ; 199(10): 1142-1150, 2023 Jun 23.
Artigo em Inglês | MEDLINE | ID: mdl-37220643

RESUMO

This study reports results of the first nationwide survey of computed tomography (CT) practice in Ukraine and proposed national diagnostic reference levels (DRLs) for common CT examinations. The collected data included characteristics of CT scanners, frequency of CT exams per anatomical region, dose indices CTDIvol and dose-length product (DLP). National DRLs were proposed at the 75th percentile of median dose indices distributions for four common CT protocols: head without contrast (brain examinations for stroke and trauma), routine chest without contrast, single-phase contrast-enhanced CT of abdomen and pelvis and oncology protocol (chest-abdomen-pelvis). The results showed 4- to 9-fold variations of median dose indices between CT scanners for the same type of examination. The following values of the CTDIvol and DLP were proposed as national DRLs: 59 mGy and 1130 mGy•cm for head, 14 mGy and 492 mGy•cm for chest, 22 mGy and 845 mGy•cm for abdomen/pelvis and 2120 mGy•cm for oncological protocol.


Assuntos
Níveis de Referência de Diagnóstico , Tomografia Computadorizada por Raios X , Doses de Radiação , Ucrânia , Valores de Referência , Tomografia Computadorizada por Raios X/métodos , Inquéritos e Questionários
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA