RESUMO
We present a multi-modal fiber array snapshot technique (M-FAST) based on an array of 96 compact cameras placed behind a primary objective lens and a fiber bundle array. Our technique is capable of large-area, high-resolution, multi-channel video acquisition. The proposed design provides two key improvements to prior cascaded imaging system approaches: a novel optical arrangement that accommodates the use of planar camera arrays, and a new ability to acquire multi-modal image data acquisition. M-FAST is a multi-modal, scalable imaging system that can acquire snapshot dual-channel fluorescence images as well as differential phase contrast measurements over a large 6.59â mm × 9.74â mm field-of-view at 2.2-µm center full-pitch resolution.
RESUMO
This work demonstrates a multi-lens microscopic imaging system that overlaps multiple independent fields of view on a single sensor for high-efficiency automated specimen analysis. Automatic detection, classification and counting of various morphological features of interest is now a crucial component of both biomedical research and disease diagnosis. While convolutional neural networks (CNNs) have dramatically improved the accuracy of counting cells and sub-cellular features from acquired digital image data, the overall throughput is still typically hindered by the limited space-bandwidth product (SBP) of conventional microscopes. Here, we show both in simulation and experiment that overlapped imaging and co-designed analysis software can achieve accurate detection of diagnostically-relevant features for several applications, including counting of white blood cells and the malaria parasite, leading to multi-fold increase in detection and processing throughput with minimal reduction in accuracy.