Rev Sci Instrum ; 89(7): 073109, 2018 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-30068109

RESUMO

We present a system level description of a cavity-enhanced millimeter-wave spectrometer that is the first in its class to combine source and detection electronics constructed from architectures commonly deployed in the mobile phone industry and traditional pulsed Fourier transform techniques to realize a compact device capable of sensitive and specific in situ gas detections. The instrument, which has an operational bandwidth of 90-102 GHz, employs several unique components, including a custom-designed pair of millimeter-wave transmitter and heterodyne receiver integrated circuit chips constructed with 65 nm complementary metal-oxide semiconductor (CMOS) techniques. These elements are directly mated to a hybrid coupling structure that enables free-space interaction of the electronics with a small gas volume while also acting as a cavity end mirror. Instrument performance for sensing of volatile compounds is highlighted with experimental trials taken in bulk gas flows and seeded molecular beam environments.