1.
Pacing Clin Electrophysiol ; 40(11): 1291-1297, 2017 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-28940232

RESUMO

BACKGROUND: A cardiac lead with a side helix for active fixation to the coronary vein wall (Attain Stability® , Model 20066, Medtronic, Minneapolis, MN, USA) recently received CE Mark. The lead is designed to improve left ventricular (LV) placement and reduce dislodgement rates. The extractability of this active fixation LV lead has not been studied extensively. METHODS: Seventeen sheep were implanted with either an LV lead with a side helix (Model 20066, Model 20096, Medtronic) or a unipolar LV lead (Model 4193, Medtronic) as a control. Leads were extracted at approximately 26, 52, or 118 weeks. Standard extraction methodology was employed with quantitative traction up to 907 g (2 lbs.) using a locking stylet. Gross pathology and histology of the heart with particular attention to the lead tracts were performed. RESULTS: All leads were successfully removed in their entirety and required significantly less than 1 kg of traction force. The side helix disengaged from the vein as designed and resulted in no complications. No cardiac tissue was observed on any extracted lead. Gross pathology and histology were devoid of any helix-induced lesions in the vascular structures. The epicardium over the side helix was normal and the fibrotic reaction around the helix was not significantly different from that around the nonhelix portions of the study leads or the control lead. CONCLUSION: Extraction of the side helix, active fixation LV lead from the coronary veins in the sheep model is safe, without procedural complexity, and free of complications after long-term LV lead implant duration.


Assuntos
Dispositivos de Terapia de Ressincronização Cardíaca , Vasos Coronários , Ventrículos do Coração , Animais , Meios de Contraste/administração & dosagem , Angiografia Coronária , Remoção de Dispositivo , Desenho de Equipamento , Modelos Animais , Carneiro Doméstico
2.
Circ Arrhythm Electrophysiol ; 5(4): 815-20, 2012 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-22787012

RESUMO

BACKGROUND: Phrenic nerve stimulation (PNS) is a common complication of cardiac resynchronization therapy when left ventricular (LV) pacing occurs via a coronary vein. The purpose of this study was to evaluate the effects of bipolar electrode spacing on PNS and LV pacing thresholds. METHODS AND RESULTS: Electrophysiology catheters with standard (2 mm-5 mm-2 mm) or modified (1 mm-5 mm-1 mm) interelectrode spacing was, respectively, inserted in a posterior/lateral cardiac vein in a randomized order in 6 anesthetized dogs via jugular access. The phrenic nerve was dissected via a left minithoracotomy and repositioned over the vein as close as possible to one of the electrodes. The presence of PNS was verified (ie, PNS threshold <2 V at 0.5 ms in unipolar configuration). Bipolar pacing was delivered using the electrode closest to the phrenic nerve as the cathode, and multiple bipolar electrode spacing configurations were tested. During bipolar pacing, PNS threshold increased as bipolar electrode spacing was reduced (P<0.05), whereas LV pacing thresholds did not change significantly (P>0.05). Compared with a standard bipolar electrode spacing of 20 mm for LV leads, 1 and 2 mm bipolar electrode spacing resulted in a PNS threshold increase of 5.5±2.2 V (P=0.003) and 2.8±1.7 V (P<0.001), respectively. Similarly, PNS threshold increased by 6.5±3.7 V with 1 mm and by 3.8±1.9 V with 2 mm bipolar pacing (both P<0.001), compared with unipolar pacing. CONCLUSIONS: This study suggests that reducing LV bipolar electrode spacing from the standard 20 mm to 1 or 2 mm may significantly increase the PNS threshold without compromising LV pacing thresholds.


Assuntos
Dispositivos de Terapia de Ressincronização Cardíaca , Terapia de Ressincronização Cardíaca/métodos , Doenças do Sistema Nervoso Periférico/prevenção & controle , Nervo Frênico/fisiopatologia , Função Ventricular Esquerda , Potenciais de Ação , Animais , Terapia de Ressincronização Cardíaca/efeitos adversos , Dispositivos de Terapia de Ressincronização Cardíaca/efeitos adversos , Angiografia Coronária , Dissecação , Cães , Desenho de Equipamento , Modelos Animais , Doenças do Sistema Nervoso Periférico/etiologia , Doenças do Sistema Nervoso Periférico/fisiopatologia , Nervo Frênico/cirurgia , Toracotomia , Fatores de Tempo