RESUMO
BACKGROUND: Endovascular thrombectomy improves outcomes and reduces mortality for large vessel occlusion (LVO) and is time-sensitive. Computer automation may aid in the early detection of LVOs, but false values may lead to alarm desensitization. We compared Viz LVO and Rapid LVO for automated LVO detection. METHODS: Data were retrospectively extracted from Rapid LVO and Viz LVO running concurrently from January 2022 to January 2023 on CT angiography (CTA) images compared with a radiologist interpretation. We calculated diagnostic accuracy measures and performed a McNemar test to look for a difference between the algorithms' errors. We collected demographic data, comorbidities, ejection fraction (EF), and imaging features and performed a multiple logistic regression to determine if any of these variables predicted the incorrect classification of LVO on CTA. RESULTS: 360 participants were included, with 47 large vessel occlusions. Viz LVO and Rapid LVO had a specificity of 0.96 and 0.85, a sensitivity of 0.87 and 0.87, a positive predictive value of 0.75 and 0.46, and a negative predictive value of 0.98 and 0.97, respectively. A McNemar test on correct and incorrect classifications showed a statistically significant difference between the two algorithms' errors (P=0.00000031). A multiple logistic regression showed that low EF (Viz P=0.00125, Rapid P=0.0286) and Modified Woodcock Score >1 (Viz P=0.000198, Rapid P=0.000000975) were significant predictors of incorrect classification. CONCLUSION: Rapid LVO produced a significantly larger number of false positive values that may contribute to alarm desensitization, leading to missed alarms or delayed responses. EF and intracranial atherosclerosis were significant predictors of incorrect predictions.