RESUMO
This article expresses the current view of the European Society of Gastrointestinal Endoscopy (ESGE) about radiation protection for endoscopic procedures, in particular endoscopic retrograde cholangiopancreatography (ERCP). Particular cases, including pregnant women and pediatric patients, are also discussed. This Guideline was developed by a group of endoscopists and medical physicists to ensure that all aspects of radiation protection are adequately dealt with. A two-page executive summary of evidence statements and recommendations is provided. The target readership for this Guideline mostly includes endoscopists, anesthesiologists, and endoscopy assistants who may be exposed to X-rays during endoscopic procedures.
Assuntos
Colangiopancreatografia Retrógrada Endoscópica/instrumentação , Colangiopancreatografia Retrógrada Endoscópica/normas , Exposição Ocupacional/análise , Segurança do Paciente/normas , Monitoramento de Radiação/normas , Proteção Radiológica/normas , Adulto , Criança , Colangiopancreatografia por Ressonância Magnética , Endossonografia , Feminino , Filtração , Fluoroscopia/métodos , Fluoroscopia/normas , Pessoal de Saúde , Humanos , Educação de Pacientes como Assunto , Gravidez , Complicações na Gravidez/diagnóstico , Garantia da Qualidade dos Cuidados de Saúde/legislação & jurisprudência , Doses de Radiação , Monitoramento de Radiação/métodos , SuíçaRESUMO
The standard Fourier metrics used for assessing image quality in computed tomography (CT) use a planar (2D) formulation of the noise equivalent quanta (NEQ) without a specific concept to manage the influence of longitudinal resolution variations, thus create a bias in the comparison of image quality for different reconstructed slice thicknesses. For this work, we designed a 'slice NEQ' that takes the longitudinal resolution into account and provides a volumetric indication of the scanner imaging performance. We also developed a specific formulation for the system DQE at the CT isocentre. A cylindrical water phantom with three different inserts for three contrast levels (-100, 120 and 1000 HU) was used for the task-based transfer functions (TTF) and noise power spectra (NPS) measurements. The longitudinal TTF was measured using the point source of the Catphan® 600 phantom. Images of the phantoms were acquired on two scanners (GE Discovery 750 HD and Siemens SOMATOM Force) and reconstructed using different slice thicknesses between 1 and 5 mm and algorithms (FBP for both systems, ASIR 50 and ASIR-V 50 for the GE). The slice NEQ correctly compared the imaging performance for different longitudinal resolutions whereas the 2D NEQ increased proportionally with the reconstructed slice thickness. The system DQE peaked at 0.70 (at 0.1 mm-1) for the Siemens and at 0.50 (at 0.1 mm-1) for the GE for FBP reconstructions. The validity of these Fourier-based metrics was restricted to a limited range of contrast due to nonlinearities introduced when dealing with iterative reconstructions (IR).
Assuntos
Tomografia Computadorizada por Raios X/métodos , Algoritmos , Humanos , Imagens de Fantasmas , Doses de Radiação , Razão Sinal-Ruído , ÁguaRESUMO
A rigorous 2D analysis of signal and noise transfer applied to reconstructed planes in digital breast tomosynthesis (DBT) is necessary for system characterization and optimization. This work proposes a method for assessing technical image quality and system detective quantum efficiency (DQEsys) for reconstructed planes in DBT. Measurements of 2D in-plane modulation transfer function (MTF) and noise power spectrum (NPS) were made on five DBT systems using different acquisition parameters, reconstruction algorithms and plane spacing. This work develops the noise equivalent quanta (NEQ), DQEsys and detectability index (d') calculated using a non-prewhitening model observer with eye filter (NPWE) for reconstructed DBT planes. The images required for this implementation were acquired using a homogeneous test object of thickness 40 mm poly(methyl) methacrylate plus 0.5 mm Al; 2D MTF was calculated from an Al disc of thickness 0.2 mm and diameter 50 mm positioned within the phantom. The radiant contrast of the MTF disc and the air kerma at the system input were used as normalization factors. The NPWE detectability index was then compared to the in-plane contrast-detail (c-d) threshold measured using the CDMAM phantom. The MTF and NPS measured on the different systems showed a strong anisotropy, consistent with the cascaded models developed in the literature for DBT. Detectability indices calculated from the measured MTF and NPS successfully predicted changes in c-d detectability for details between 0.1 mm and 2.0 mm, for DBT plane spacings between 0.5 mm and 10 mm, and for air kerma values at the system input between 157 µGy and 1170 µGy. The linear Pearson correlation between the detectability index and threshold gold thickness of the CDMAM phantom was -0.996. The method implements a parametric means of assessing the technical image quality of reconstructed DBT planes, providing valuable information for optimization of DBT systems.
Assuntos
Algoritmos , Mama/diagnóstico por imagem , Mamografia/métodos , Imagens de Fantasmas , Controle de Qualidade , Intensificação de Imagem Radiográfica/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/normas , Feminino , HumanosRESUMO
We used the Timepix3 hybrid pixel detector technology in order to determine the exposure of medical personnel to ionizing radiation in an interventional radiology room. We measured the energy spectra of the scattered radiation generated by the patient during X-ray image-guided interventional procedures. We performed measurements at different positions and heights within the theatre. We first observed a difference in fluence for each staff member. As expected, we found that the person closest to the X-ray tube is the most exposed while the least exposed staff member is positioned at the patient's feet. Additionally, we observed a shift in energy from head to toe for practitioners, clearly indicating a non-homogenous energy exposure. The photon counting Timepix3 detector provides a new tool for radiation field characterisation that is easier-to-use and more compact than conventional X-ray spectrometers. The spectral information is particularly valuable for optimising the use of radiation protection gear and improving dosimetry surveillance programs. We also found the device very useful for training purposes to provide awareness and understanding about radiation protection principles among interventional radiology staff.
Assuntos
Radiologia Intervencionista/métodos , Humanos , Exposição Ocupacional , Doses de Radiação , Lesões por Radiação/prevenção & controle , Espalhamento de RadiaçãoRESUMO
The standard approach to signal difference-to-noise ratio (SDNR) analysis requires a region of interest (ROI) positioned within the object to measure signal-difference, restricting this metric to flat-topped objects with large, sharply delineated areas. This work develops a generalized expression for SDNR (SDNRg) calculated from a ROI encompassing the object. Signal power, defined as the deviation of pixel values from the mean background due to the object, is used instead of signal-difference. Comparison was first made by simulating ideal flat-topped discs with sharp edges and diameters between 1 and 80â¯pixels, into a uniformly noisy background using a known signal-difference. For discs covering more than 20â¯pixels, SDNRg and standard SDNR (SDNRst) were within 3%, while for discs of less than 20â¯pixels, SDNRg was within 26% of the truth compared to 58% for SDNRst. Generalized and standard SDNR were compared for radiography images of three different phantoms with microcalcification-like objects (MTM-100 phantom), hemispheric objects of different thicknesses with a Gaussian intensity distribution and mammography quality control (QC) images. Applied to Gaussian details, SDNRg was between 20% and 45% higher than SDNRst, depending on object thickness, while for the QC images, SDNRg was with 1.7% of the standard SDNR. Compared to the standard SDNR, SDNRg is applicable to non-uniform signals, where an explicit contrast measurement is not suitable, and has improved accuracy when assessing SDNR of small objects.
Assuntos
Processamento de Imagem Assistida por Computador/métodos , Mamografia , Razão Sinal-RuídoRESUMO
An objective analysis of image quality parameters was performed for six digital mammography systems. The presampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) for the systems were determined at different doses, for 28 kVp with a Mo/Mo or W/Al target/filter combination and 2 mm of additional aluminium filtration. The flat-panel units have higher MTF and DQE in the mid to high frequency range than standard CR systems. The highest DQE, over the whole dose range, is for the slit-scanning direct photon counting system. Dual-side read CR can overcome the inherent x-ray absorption and signal collection limitations of standard CR mammography, improving the low-frequency DQE by 40%, to the same level as full-field systems, but it does not improve the poor spatial resolution of phosphor.
Assuntos
Mamografia/instrumentação , Mamografia/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Sistemas Computacionais , Desenho de Equipamento , Feminino , Humanos , Modelos Estatísticos , Fótons , Reprodutibilidade dos Testes , Software , Tecnologia Radiológica/métodos , Ecrans Intensificadores para Raios X , Raios XRESUMO
A nationwide investigation was conducted in Switzerland to establish the exposure of the population by medical x rays and update the results of the 1998 survey. Both the frequency and the dose variations were studied in order to determine the change in the collective dose. The frequency study addressed 206 general practitioners (GPs), 30 hospitals, and 10 private radiology institutes. Except for the latter, the response rate was very satisfactory. The dose study relied on the assessment of the speed class of the screen-film combinations used by the GPs as well as the results of two separate studies dedicated to fluoroscopy and CT. The investigation showed that the total number of all medical x-ray examinations performed by GPs registered a 1% decrease between 1998 and 2003, and that the sensitivities of the film-screen combinations registered a shift towards higher values, leading to a reduction of the dose delivered by a GP of the order of 20%. The study indicated also that the total number of all x-ray examinations performed in hospitals increased by 4%, with a slight increase of radiographies by 1% but significant decrease of examinations involving fluoroscopy (39%), and a 70% increase for CT examinations. Concerning the doses, the investigation of a selection of examinations involving fluoroscopy showed a significant increase of the kerma-area product (KAP) per procedure. For CT the study showed an increase of the dose-length product (DLP) per procedure for skull and abdomen examinations, and a decrease for chest examination. Both changes in the frequency and the effective dose per examination led to a 20% increase in the total collective dose.
Assuntos
Carga Corporal (Radioterapia) , Exposição Ambiental/análise , Doses de Radiação , Monitoramento de Radiação , Radiografia/estatística & dados numéricos , Medição de Risco/métodos , Humanos , Eficiência Biológica Relativa , Fatores de Risco , Suíça/epidemiologiaRESUMO
The age of the patient is of prime importance when assessing the radiological risk to patients due to medical X-ray exposures and the total detriment to the population due to radiodiagnostics. In order to take into account the age-specific radiosensitivity, three age groups are considered: children, adults and the elderly. In this work, the relative number of examinations carried out on paediatric and geriatric patients is established, compared with adult patients, for radiodiagnostics as a whole, for dental and medical radiology, for 8 radiological modalities as well as for 40 types of X-ray examinations. The relative numbers of X-ray examinations are determined based on the corresponding age distributions of patients and that of the general population. Two broad groups of X-ray examinations may be defined. Group A comprises conventional radiography, fluoroscopy and computed tomography; for this group a paediatric patient undergoes half the number of examinations as that of an adult, and a geriatric patient undergoes 2.5 times more. Group B comprises angiography and interventional procedures; for this group a paediatric patient undergoes a one-fourth of the number of examinations carried out on an adult, and a geriatric patient undergoes five times more.
Assuntos
Radiografia/classificação , Radiografia/estatística & dados numéricos , Medição de Risco/métodos , Adolescente , Adulto , Distribuição por Idade , Idoso , Idoso de 80 Anos ou mais , Criança , Pré-Escolar , Feminino , Humanos , Lactente , Recém-Nascido , Masculino , Pessoa de Meia-Idade , Fatores de Risco , SuíçaRESUMO
This work proposes a method for assessing the detective quantum efficiency (DQE) of radiographic imaging systems that include both the x-ray detector and the antiscatter device. Cascaded linear analysis of the antiscatter device efficiency (DQEASD) with the x-ray detector DQE is used to develop a metric of system efficiency (DQEsys); the new metric is then related to the existing system efficiency parameters of effective DQE (eDQE) and generalized DQE (gDQE). The effect of scatter on signal transfer was modelled through its point spread function (PSF), leading to an x-ray beam transfer function (BTF) that multiplies with the classical presampling modulation transfer function (MTF) to give the system MTF. Expressions are then derived for the influence of scattered radiation on signal-difference to noise ratio (SDNR) and contrast-detail (c-d) detectability. The DQEsys metric was tested using two digital mammography systems, for eight x-ray beams (four with and four without scatter), matched in terms of effective energy. The model was validated through measurements of contrast, SDNR and MTF for poly(methyl)methacrylate thicknesses covering the range of scatter fractions expected in mammography. The metric also successfully predicted changes in c-d detectability for different scatter conditions. Scatter fractions for the four beams with scatter were established with the beam stop method using an extrapolation function derived from the scatter PSF, and validated through Monte Carlo (MC) simulations. Low-frequency drop of the MTF from scatter was compared to both theory and MC calculations. DQEsys successfully quantified the influence of the grid on SDNR and accurately gave the break-even object thickness at which system efficiency was improved by the grid. The DQEsys metric is proposed as an extension of current detector characterization methods to include a performance evaluation in the presence of scattered radiation, with an antiscatter device in place.
Assuntos
Modelos Teóricos , Radiografia/métodos , Espalhamento de Radiação , Método de Monte Carlo , Imagens de Fantasmas , Intensificação de Imagem Radiográfica , Razão Sinal-RuídoRESUMO
In our recent paper by Monnin et al. [Med. Phys. 33, 411-420 (2006)], an objective analysis of the relative performance of a computed radiography (CR) system using both standard single-side (ST-VI) and prototype dual-side read (ST-BD) plates was reported. The presampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of paediatric chest radiography, at an entrance detector air kerma of 5 microGy. Experiments demonstrated that, compared to the standard single-side read system, the MTF for the dual-side read system was slightly reduced, but a significant decrease in image noise resulted in a marked increase in DQE (+40%) in the low spatial frequency range. However, the DQE improvement for the ST-BD plate decreased with increasing spatial frequency, and, at spatial frequencies above 2.2 mm(-1), the DQE of the dual-side read system was lower than that of the single-side one.
Assuntos
Intensificação de Imagem Radiográfica/instrumentação , Interpretação de Imagem Radiográfica Assistida por Computador/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Humanos , Lasers , Radiometria , Reprodutibilidade dos Testes , Espalhamento de Radiação , Sensibilidade e Especificidade , Espectrometria por Raios XRESUMO
An objective analysis of image quality parameters was performed for a computed radiography (CR) system using both standard single-side and prototype dual-side read plates. The pre-sampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of pediatric chest radiography, at an entrance detector air kerma of 5 microGy. The NPS and DQE measurements were realized under clinically relevant x-ray spectra for pediatric radiology, including x-ray scatter radiations. Compared to the standard single-side read system, the MTF for the dual-side read system is reduced, but this is offset by a significant decrease in image noise, resulting in a marked increase in DQE (+40%) in the low spatial frequency range. Thus, for the same image quality, the new technology permits the CR system to be used at a reduced dose level.
Assuntos
Imagens de Fantasmas , Intensificação de Imagem Radiográfica/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Criança , Análise de Falha de Equipamento , Humanos , Intensificação de Imagem Radiográfica/instrumentação , Interpretação de Imagem Radiográfica Assistida por Computador/instrumentação , Radiometria , Reprodutibilidade dos Testes , Espalhamento de Radiação , Sensibilidade e Especificidade , Espectrometria por Raios XRESUMO
PURPOSE: This study analyzes CT examinations in Switzerland. MATERIALS AND METHODS: Using different sources (administrative data on the equipment, a 1998 nationwide inquiry into practices, and data provided by the Swiss University Hospitals of Basel, Zurich, and Lausanne), we determined the frequency of CT examinations (hospitals and private radiologists) in 1998 according to different descriptive variables and studied the progression in CT use over time. RESULTS: CT scanners increased by 7% between 1998 and 2004. The average annual number of CT examinations in 1998 was 46.3/1000 population, 3.4% of all radiological examinations in Switzerland in 1997-1998. The most frequent examination was CT of the skull (24%), while private radiology institutes perform more CTs of the spine. More CT examinations were performed for men than for women (sex ratio M/F=1.17). The average annual increase in CT in Swiss hospitals varied from 8% for Basel to 18% for Lausanne. Finally, the proportion of pediatric examinations was 5%; their numbers appear to be stabilizing. CONCLUSION: There is a significant increase in CT examinations. It is hoped that our study will heighten awareness among doctors of CT examinations in order to optimize their use.
Assuntos
Tomografia Computadorizada por Raios X/estatística & dados numéricos , Tomografia Computadorizada por Raios X/tendências , Humanos , Suíça , Fatores de TempoRESUMO
The development of CT applications might become a public health problem if no effort is made on the justification and the optimisation of the examinations. This paper presents some hints to assure that the risk-benefit compromise remains in favour of the patient, especially when one deals with the examinations of young patients. In this context a particular attention has to be made on the justification of the examination. When performing the acquisition one needs to optimise the extension of the volume investigated together with the number of acquisition sequences used. Finally, the use of automatic exposure systems, now available on all the units, and the use of the Diagnostic Reference Levels (DRL) should allow help radiologists to control the exposure of their patients.
Assuntos
Doses de Radiação , Tomografia Computadorizada por Raios X/métodos , Criança , Humanos , Lesões por Radiação/prevenção & controleRESUMO
A version of cascaded systems analysis was developed specifically with the aim of studying quantum noise propagation in x-ray detectors. Signal and quantum noise propagation was then modelled in four types of x-ray detectors used for digital mammography: four flat panel systems, one computed radiography and one slot-scan silicon wafer based photon counting device. As required inputs to the model, the two dimensional (2D) modulation transfer function (MTF), noise power spectra (NPS) and detective quantum efficiency (DQE) were measured for six mammography systems that utilized these different detectors. A new method to reconstruct anisotropic 2D presampling MTF matrices from 1D radial MTFs measured along different angular directions across the detector is described; an image of a sharp, circular disc was used for this purpose. The effective pixel fill factor for the FP systems was determined from the axial 1D presampling MTFs measured with a square sharp edge along the two orthogonal directions of the pixel lattice. Expectation MTFs were then calculated by averaging the radial MTFs over all possible phases and the 2D EMTF formed with the same reconstruction technique used for the 2D presampling MTF. The quantum NPS was then established by noise decomposition from homogenous images acquired as a function of detector air kerma. This was further decomposed into the correlated and uncorrelated quantum components by fitting the radially averaged quantum NPS with the radially averaged EMTF(2). This whole procedure allowed a detailed analysis of the influence of aliasing, signal and noise decorrelation, x-ray capture efficiency and global secondary gain on NPS and detector DQE. The influence of noise statistics, pixel fill factor and additional electronic and fixed pattern noises on the DQE was also studied. The 2D cascaded model and decompositions performed on the acquired images also enlightened the observed quantum NPS and DQE anisotropy.
Assuntos
Mamografia/métodos , Intensificação de Imagem Radiográfica/métodos , Mamografia/instrumentação , Mamografia/normas , Intensificação de Imagem Radiográfica/instrumentação , Intensificação de Imagem Radiográfica/normas , Razão Sinal-RuídoRESUMO
As opposed to the standard detective quantum efficiency (DQE), effective DQE (eDQE) is a figure of merit that allows comparing the performances of imaging systems in the presence of scatter rejection devices. The geometry of the EOS™ slot-scanning system is such that the detector is self-collimated and rejects scattered radiation. In this study, the EOS system was characterised using the eDQE in imaging conditions similar to those used in clinical practice: with phantoms of different widths placed in the X-ray beam, for various incident air kerma and tube voltages corresponding to the phantom thickness. Scatter fractions in EOS images were extremely low, around 2 % for all configurations. Maximum eDQE values spanned 9-14.8 % for a large range of air kerma at the detector plane from 0.01 to 1.34 µGy. These figures were obtained with non-optimised EOS setting but still over-performed most of the maximum eDQEs recently assessed for various computed radiology and digital radiology systems with antiscatter grids.
Assuntos
Imageamento Tridimensional/instrumentação , Intensificação de Imagem Radiográfica/instrumentação , Interpretação de Imagem Radiográfica Assistida por Computador/instrumentação , Tomografia Computadorizada por Raios X/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Imagens de Fantasmas , Doses de Radiação , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , TransdutoresRESUMO
Four standard radiation qualities (from RQA 3 to RQA 9) were used to compare the imaging performance of a computed radiography (CR) system (general purpose and high resolution phosphor plates of a Kodak CR 9000 system), a selenium-based direct flat panel detector (Kodak Direct View DR 9000), and a conventional screen-film system (Kodak T-MAT L/RA film with a 3M Trimax Regular screen of speed 400) in conventional radiography. Reference exposure levels were chosen according to the manufacturer's recommendations to be representative of clinical practice (exposure index of 1700 for digital systems and a film optical density of 1.4). With the exception of the RQA 3 beam quality, the exposure levels needed to produce a mean digital signal of 1700 were higher than those needed to obtain a mean film optical density of 1.4. In spite of intense developments in the field of digital detectors, screen-film systems are still very efficient detectors for most of the beam qualities used in radiology. An important outcome of this study is the behavior of the detective quantum efficiency of the digital radiography (DR) system as a function of beam energy. The practice of users to increase beam energy when switching from a screen-film system to a CR system, in order to improve the compromise between patient dose and image quality, might not be appropriate when switching from screen-film to selenium-based DR systems.
Assuntos
Intensificação de Imagem Radiográfica/instrumentação , Interpretação de Imagem Radiográfica Assistida por Computador/instrumentação , Tomografia Computadorizada por Raios X/instrumentação , Ecrans Intensificadores para Raios X , Reprodutibilidade dos Testes , Sensibilidade e EspecificidadeRESUMO
Three standard radiation qualities (RQA 3, RQA 5 and RQA 9) and two screens, Kodak Lanex Regular and Insight Skeletal, were used to compare the imaging performance and dose requirements of the new Kodak Hyper Speed G and the current Kodak T-MAT G/RA medical x-ray films. The noise equivalent quanta (NEQ) and detective quantum efficiencies (DQE) of the four screen-film combinations were measured at three gross optical densities and compared with the characteristics for the Kodak CR 9000 system with GP (general purpose) and HR (high resolution) phosphor plates. The new Hyper Speed G film has double the intrinsic sensitivity of the T-MAT G/RA film and a higher contrast in the high optical density range for comparable exposure latitude. By providing both high sensitivity and high spatial resolution, the new film significantly improves the compromise between dose and image quality. As expected, the new film has a higher noise level and a lower signal-to-noise ratio than the standard film, although in the high frequency range this is compensated for by a better resolution, giving better DQE results--especially at high optical density. Both screen-film systems outperform the phosphor plates in terms of MTF and DQE for standard imaging conditions (Regular screen at RQA 5 and RQA 9 beam qualities). At low energy (RQA 3), the CR system has a comparable low-frequency DQE to screen-film systems when used with a fine screen at low and middle optical densities, and a superior low-frequency DQE at high optical density.
Assuntos
Interpretação de Imagem Radiográfica Assistida por Computador , Filme para Raios X , Ecrans Intensificadores para Raios XRESUMO
This work compares the detector performance and image quality of the new Kodak Min-R EV mammography screen-film system with the Fuji CR Profect detector and with other current mammography screen-film systems from Agfa, Fuji and Kodak. Basic image quality parameters (MTF, NPS, NEQ and DQE) were evaluated for a 28 kV Mo/Mo (HVL = 0.646 mm Al) beam using different mAs exposure settings. Compared with other screen-film systems, the new Kodak Min-R EV detector has the highest contrast and a low intrinsic noise level, giving better NEQ and DQE results, especially at high optical density. Thus, the properties of the new mammography film approach those of a fine mammography detector, especially at low frequency range. Screen-film systems provide the best resolution. The presampling MTF of the digital detector has a value of 15% at the Nyquist frequency and, due to the spread size of the laser beam, the use of a smaller pixel size would not permit a significant improvement of the detector resolution. The dual collection reading technology increases significantly the low frequency DQE of the Fuji CR system that can at present compete with the most efficient mammography screen-film systems.
Assuntos
Mamografia/instrumentação , Algoritmos , Processamento de Imagem Assistida por Computador , Mamografia/métodos , Imagens de Fantasmas , Software , Filme para Raios XRESUMO
The purpose of this work was to compare standard desktop display systems with dedicated medical display systems. The set of image tests proposed by the American Association of Physicists in Medicine (AAPM TG18) was used to assess a Philips 107S desktop display system and a Siemens medical display. Three observers performed the subjective assessment, in a non-concerted manner. The objective assessment was performed using a CCD camera according to the AAPM TG18 procedure. The results clearly demonstrate the inadequacy of standard desktop display systems in the framework of diagnostic radiology. Moreover, a good correlation between the subjective and objective assessment methods was obtained.
Assuntos
Diagnóstico por Imagem/instrumentação , Radiografia/métodos , Radiografia/normas , Radiologia/métodos , Radiometria/métodos , Radiometria/normas , Terminais de Computador , Computadores , Apresentação de Dados , Diagnóstico por Imagem/métodos , Humanos , Luz , Intensificação de Imagem Radiográfica , Sistemas de Informação em Radiologia , Sensibilidade e Especificidade , Software , Interface Usuário-ComputadorRESUMO
A high-end drum film digitiser (Tango, Germany) and a calibrated linear microdensitometer developed by PTB were used to assess the modulation transfer function (MTF) and the noise power spectra (NPS) of 3 mammographic screen film systems at optical density levels of 0.8, 1.5 and 2.5. The use of a drum scanner to assess MTF and NPS data appears to be adequate but requires an appropriate characterisation of the scanner to verify its internal noise level and its MTF. It is further necessary to calibrate the scanner output in terms of visual diffuse optical densities. Processing of two-dimensional digital data of grating images need to be more strictly defined for accurate MTF measurements of screen-film systems. Nevertheless, even now it seems to be feasible to use commercially available high-end and well calibrated scanners to assess screen film systems. This is especially important for quality assurance purposes because important parameters of screen film systems such like MTF and NPS can now be determined without using sophisticated microdensitometers which are not commercially available.