Anal Bioanal Chem ; 373(6): 356-67, 2002 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-12172670

RESUMO

More than a decade after Phillips' first published work this article reviews recent developments in comprehensive two-dimensional gas chromatography (GC x GC). Special attention is devoted to the further development and diversity of modulation devices. These include heated sweepers, cryofocused modulators, and a variety of diaphragm valve-switching strategies. It is demonstrated that all modulation approaches can be very well suited to GC x GC, depending on the particular application. Diaphragm-valve modulation is very powerful for volatile organic compounds. Slotted heater and cryofocused modulation are preferred for samples that contain non-volatile components. Applications ranging from petroleum to environmental and biological samples are illustrated. Extension of the technique to GC x GC-mass spectrometry (MS) is also discussed and trends for future research activity are pointed out.


Assuntos
Cromatografia Gasosa/métodos , Cromatografia Gasosa/instrumentação , Desenho de Equipamento , Óleos Voláteis/análise , Praguicidas/análise , Sensibilidade e Especificidade , Poluentes Químicos da Água/análise