Acta Radiol ; : 2841851241269918, 2024 Aug 20.
Artigo em Inglês | MEDLINE | ID: mdl-39161325

RESUMO

BACKGROUND: High-resolution computed tomography (HRCT) is dependent on detailed morphology in diagnostic assessment of interstitial lung diseases. Photon-counting CT (PCCT) enables improved resolution while reducing radiation. PURPOSE: To compare if the image quality, detailed morphology, and radiation dose in HRCT of the lung improves with PCCT compared to energy-integrated CT (EICT). MATERIAL AND METHODS: HRCT with PCCT in patients with body mass index (BMI) from normal to obese, previously examined with different EICT were included. They were evaluated in a five-step scale for image quality according to Quality Criteria for CT (Diagnostic Requirement of the ImPACT group-European standardization). In addition, ground-glass opacities, bronchiectasis, emphysema, nodules, and subpleural detailed morphology (≤1 cm from the pleural border) were evaluated by three independent thoracic and/or pediatric radiologists. Visual grading characteristics (VGC) were used for comparison of image quality and detailed morphology and Fleiss kappa for intra-observer variability. Dose-length product (DLP) and CT dose index-volume (CTDIvol) were collected to calculate effective radiation dose. RESULTS: HRCT with PCCT in 52 women and 48 men (mean age=67.2 ± 13.6 years; age range=27-87 years; BMI=26.9 kg/m2; range=18.6-45 kg/m2) previously examined with EICT (mean age=65.3 ± 13.6 years; age range=27-85 years; BMI=27 kg/m2; range=18.9-45 kg/m2) were included. There were significant differences in image quality for all entities in favor of PCCT. The radiation dose was reduced with PCCT by 47% in all, particularly pronounced in obese with 48.5%. CONCLUSION: Image quality, detailed morphology, and radiation dose, particularly in obese patients, were significantly improved in HRCT with PCCT compared to conventional EICT. The new technique enables visualization of subpleural structures.

2.
Phys Med ; 31(2): 173-8, 2015 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-25555906

RESUMO

PURPOSE: To develop methods for qualitative and quantitative evaluation of MRI artifacts near metallic prostheses, and to compare the efficiency of different artifact suppression techniques with different types of hip prostheses. METHODS: Three hip prostheses of cobalt-chromium, stainless steel, and titanium were embedded in agarose gel together with a rectilinear grid. Coronal MR images of the prostheses were acquired on a 1.5T scanner. Three pulse sequences were evaluated; TSE: a high-bandwidth turbo spin echo; VAT: TSE with view angle tilting, SEMAC: TSE with both VAT and slice distortion correction (6, 10 or 16 z-phase-encoding steps). Through-plane distortions were assessed as the length of visible gridlines, in-plane artifacts as the artifact area, and total artifacts by subtraction of an ideal, undistorted image from the actual image. RESULTS: VAT reduced in-plane artifacts by up to 50% compared to TSE, but did not reduce through-plane artifacts. SEMAC reduced through-plane artifacts by 60-80% compared to TSE and VAT. SEMAC in-plane artifacts were from 20% higher (6 encoding steps) to 50% lower (16 steps) than VAT. Total artifacts were reduced by 60-80% in the best sequence (SEMAC, 16 steps) compared to the worst (TSE). The titanium prosthesis produced 3-4 times lower artifact scores than the other prostheses. CONCLUSIONS: A rectilinear grid phantom is useful for qualitative and quantitative evaluation of artifacts provoked by different MRI protocols and prosthesis models. VAT and SEMAC were superior to TSE with high bandwidth. A proper number of z-encoding steps in SEMAC was critical. The titanium prosthesis caused least artifacts.


Assuntos
Artefatos , Prótese de Quadril , Imageamento por Ressonância Magnética/instrumentação , Metais , Imagens de Fantasmas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA