Your browser doesn't support javascript.
loading
Measuring energy expenditure in the intensive care unit: a comparison of indirect calorimetry by E-sCOVX and Quark RMR with Deltatrac II in mechanically ventilated critically ill patients.
Rehal, Martin Sundström; Fiskaare, Erik; Tjäder, Inga; Norberg, Åke; Rooyackers, Olav; Wernerman, Jan.
Affiliation
  • Rehal MS; Department of Anesthesiology and Intensive Care Medicine, K32, Karolinska University Hospital Huddinge, Hälsovägen 13, 14186, Stockholm, Sweden. martin.sundstrom@gmail.com.
  • Fiskaare E; Division of Anesthesia and Intensive Care, Department of Clinical Sciences, Intervention and Technology (CLINTEC), Karolinska Institutet, Hälsovägen 13, 14186, Stockholm, Sweden. martin.sundstrom@gmail.com.
  • Tjäder I; Division of Anesthesia and Intensive Care, Department of Clinical Sciences, Intervention and Technology (CLINTEC), Karolinska Institutet, Hälsovägen 13, 14186, Stockholm, Sweden. erik_fiskaare85@hotmail.com.
  • Norberg Å; Department of Anesthesiology and Intensive Care Medicine, K32, Karolinska University Hospital Huddinge, Hälsovägen 13, 14186, Stockholm, Sweden. inga.tjader@karolinska.se.
  • Rooyackers O; Division of Anesthesia and Intensive Care, Department of Clinical Sciences, Intervention and Technology (CLINTEC), Karolinska Institutet, Hälsovägen 13, 14186, Stockholm, Sweden. inga.tjader@karolinska.se.
  • Wernerman J; Department of Anesthesiology and Intensive Care Medicine, K32, Karolinska University Hospital Huddinge, Hälsovägen 13, 14186, Stockholm, Sweden. ake.norberg@karolinska.se.
Crit Care ; 20: 54, 2016 Mar 05.
Article in En | MEDLINE | ID: mdl-26951095
ABSTRACT

BACKGROUND:

Indirect calorimetry allows the determination of energy expenditure in critically ill patients by measuring oxygen consumption (VO2) and carbon dioxide production (VCO2). Recent studies have demonstrated variable performance of "breath-by-breath" instruments compared to mixing chamber technology. The aim of this study was to validate two modern devices (E-sCOVX and Quark RMR) against a reference method (Deltatrac II).

METHOD:

Measurements of VO2/VCO2 with the test and reference devices were performed simultaneously over a 20-min period in mechanically ventilated adult intensive care unit patients. Accuracy and precision of instruments were analyzed using Bland-Altman plots.

RESULTS:

Forty-eight measurements in 22 patients were included for analysis. Both E-sCOVX and Quark RMR overestimated VO2 and VCO2 compared to Deltatrac II, corresponding to a 10% higher mean resting energy expenditure. Limits of agreement of resting energy expenditure within ± 2 standard deviations were ± 461 kcal/24 h (± 21% expressed as percentage error) for ΔE-sCOVX-Deltatrac II and ± 465 kcal/24 h (± 22%) for ΔQuark RMR-Deltatrac II.

CONCLUSION:

Both test devices overestimate VO2 and VCO2 compared to Deltatrac II. The observed limits of agreement are comparable to those commonly accepted in evaluations of circulatory monitoring, and significantly less than results from predictive equations. We hypothesize that the discrepancy between methods is due to patient/ventilator-related factors that affect the synchronization of gas and spirometry waveforms. TRIAL REGISTRATION Australian New Zealand Clinical Trials Registry, Trial ID ACTRN12615000205538. Date registered 3 March 2015.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Respiration, Artificial / Calorimetry, Indirect / Critical Illness / Energy Metabolism / Monitoring, Physiologic Type of study: Health_economic_evaluation / Prognostic_studies Limits: Adult / Aged / Female / Humans / Male / Middle aged Country/Region as subject: Oceania Language: En Journal: Crit Care Year: 2016 Type: Article Affiliation country: Sweden

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Respiration, Artificial / Calorimetry, Indirect / Critical Illness / Energy Metabolism / Monitoring, Physiologic Type of study: Health_economic_evaluation / Prognostic_studies Limits: Adult / Aged / Female / Humans / Male / Middle aged Country/Region as subject: Oceania Language: En Journal: Crit Care Year: 2016 Type: Article Affiliation country: Sweden