The association between defibrillation shock energy and acute cardiac damage in patients with implantable cardioverter defibrillators.
J Arrhythm
; 32(6): 481-485, 2016 Dec.
Article
in En
| MEDLINE
| ID: mdl-27920833
ABSTRACT
BACKGROUND:
The aim of this study was to establish a minimally invasive defibrillation testing (DT) protocol for patients with implantable cardioverter defibrillators (ICDs).METHODS:
Two different energy DTs were performed, immediately after (15 J-DT) and 7 days after (≤10 J-DT) device implantation, in 20 consecutive ICD implantation patients. Cardiac-troponin T (c-TNT) and heart-type fatty acid binding protein (H-FABP) levels were measured before implantation, 2 h after implantation, and 1 day after each DT. For an additional 122 patients with ICD, we retrospectively analyzed 203 DTs immediately and 7 days after device implantation.RESULTS:
Serum c-TNT levels were significantly elevated 2 h after 15 J-DT [0.008 (0.004-0.019) vs. 0.053 (0.037-0.068) ng/mL, p<0.001], but not ≤10 J-DT [0.007 (0.004-0.018) ng/mL]. Similarly, serum H-FABP levels were significantly elevated 2 h after 15 J-DT (2.9±1.5 vs. 6.4±3.4 ng/mL, p<0.001), but not ≤10 J-DT (2.7±1.5 ng/mL). The changes in c-TNT and H-FABP levels between baseline and 2 h after DT were significantly greater for 15 J-DT compared with ≤10 J-DT [c-TnT 0.039 (0.029-0.060) vs. 0 (0-0.003) ng/mL, p<0.001; H-FABP 3.6±2.8 vs. -0.16±1.1 ng/mL, p<0.001]. The success rates of the initial shocks delivered for ventricular fibrillation were no different between ≤10 J-DT (85% [78/92]) and ≥15 J-DT (92% [103/111]).CONCLUSIONS:
Elevated levels of myocardial damage markers such as c-TNT and H-FABP were not found after ≤10 J-DT. In addition, an acceptable success rate was confirmed in ≤10 J-DT.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Type of study:
Risk_factors_studies
Language:
En
Journal:
J Arrhythm
Year:
2016
Type:
Article
Affiliation country:
Japan