Your browser doesn't support javascript.
loading
Evaluation of the Gore TAG thoracic branch endoprosthesis in the treatment of proximal descending thoracic aortic aneurysms.
Dake, Michael D; Fischbein, Michael P; Bavaria, Joseph E; Desai, Nimesh D; Oderich, Gustavo; Singh, Michael J; Fillinger, Mark; Suckow, Bjoern D; Matsumura, Jon S; Patel, Himanshu J.
Affiliation
  • Dake MD; Department of Medical Imaging, University of Arizona Health System, Tuscon, Ariz. Electronic address: mddake@email.arizona.edu.
  • Fischbein MP; Department of Cardiothoracic Surgery, Stanford University Hospitals, Palo Alto, Calif.
  • Bavaria JE; Department of Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pa.
  • Desai ND; Department of Surgery, Hospital of the University of Pennsylvania, Philadelphia, Pa.
  • Oderich G; Department of Surgery, Mayo Clinic, Rochester, Minn.
  • Singh MJ; Department of Surgery, University of Pittsburgh Medical Center, Pittsburgh, Pa.
  • Fillinger M; Department of Surgery, Dartmouth-Hitchcock Medical Center, Lebanon, NH.
  • Suckow BD; Department of Surgery, Dartmouth-Hitchcock Medical Center, Lebanon, NH.
  • Matsumura JS; Department of Surgery, University of Wisconsin School of Medicine and Public Health, Madison, Wis.
  • Patel HJ; Department of Cardiac Surgery, University of Michigan Frankel Cardiovascular Center, Ann Arbor, Mich.
J Vasc Surg ; 74(5): 1483-1490.e2, 2021 11.
Article in En | MEDLINE | ID: mdl-33940079
ABSTRACT

BACKGROUND:

Thoracic endovascular aortic repair has radically transformed the treatment of descending thoracic aortic aneurysms. However, when aneurysms involve the aortic arch in the region of the left subclavian artery, branch vessel preservation must be considered. Branched aortic endografts have provided a new option to maintain branch patency.

METHODS:

Six investigative sites enrolled 31 patients in a nonrandomized, prospective investigational device exemption feasibility trial of a single branched aortic endograft for the management of aneurysms that include the distal aortic arch. The Gore TAG thoracic branch endoprosthesis (W. L. Gore & Associates, Inc, Flagstaff, Ariz), an investigational device, allows for graft placement proximal to the left subclavian artery and incorporates a single side branch for left subclavian perfusion.

RESULTS:

All 31 patients (100%) had undergone successful implantation of the investigational device in landing zone 2. Men slightly outnumbered women (51.6%). Their average age was 74.1 ± 10.4 years. The aneurysm morphology was fusiform in 12 and saccular in 19 patients, with a mean maximum aortic diameter of 54.8 ± 10.9 mm. The mean follow-up period for the cohort was 25.2 ± 11.1 months. We have reported the patient outcomes at 1 month and 1 year. At 1 month, the side branch patency was 100% and the freedom from core laboratory-reported device-related endoleak (types I and III) was 96.7%, without 30-day death or permanent paraplegia. One patient experienced a procedure-related stroke. Through 1 year, five patients had died; none of the deaths were related to the device or procedure (clinical endpoint committee adjudicated). One thoracic reintervention was required. No conversions were required, and no aneurysm growth (core laboratory) was reported. One case of the loss of side branch patency was diagnosed in the left subclavian artery in an asymptomatic individual from computed tomography at 6 months, with no reported subsequent adverse events due to loss of patency. Endoleaks were reported by the core laboratory in five patients at 12 months (two, type II; and three, indeterminate).

CONCLUSIONS:

The present investigational device exemption feasibility study has reported the preliminary results of the use of a single side branch endograft to treat patients with proximal descending thoracic aortic aneurysms.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Blood Vessel Prosthesis / Stents / Aortic Aneurysm, Thoracic / Blood Vessel Prosthesis Implantation / Endovascular Procedures Type of study: Clinical_trials / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male / Middle aged Country/Region as subject: America do norte Language: En Journal: J Vasc Surg Journal subject: ANGIOLOGIA Year: 2021 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Blood Vessel Prosthesis / Stents / Aortic Aneurysm, Thoracic / Blood Vessel Prosthesis Implantation / Endovascular Procedures Type of study: Clinical_trials / Observational_studies / Risk_factors_studies Limits: Aged / Aged80 / Female / Humans / Male / Middle aged Country/Region as subject: America do norte Language: En Journal: J Vasc Surg Journal subject: ANGIOLOGIA Year: 2021 Type: Article