Your browser doesn't support javascript.
loading
Temperature dependence, accuracy, and repeatability of T1 and T2 relaxation times for the ISMRM/NIST system phantom measured using MR fingerprinting.
Statton, Ben K; Smith, Joely; Finnegan, Mary E; Koerzdoerfer, Gregor; Quest, Rebecca A; Grech-Sollars, Matthew.
Affiliation
  • Statton BK; Medical Research Council, London Institute of Medical Sciences, Imperial College London, London, United Kingdom.
  • Smith J; Department of Imaging, Imperial College Healthcare NHS Trust, London, United Kingdom.
  • Finnegan ME; Department of Bioengineering, Imperial College London, London, United Kingdom.
  • Koerzdoerfer G; Department of Imaging, Imperial College Healthcare NHS Trust, London, United Kingdom.
  • Quest RA; Department of Bioengineering, Imperial College London, London, United Kingdom.
  • Grech-Sollars M; Siemens Healthcare GmbH, Erlangen, Germany.
Magn Reson Med ; 87(3): 1446-1460, 2022 03.
Article in En | MEDLINE | ID: mdl-34752644
ABSTRACT

PURPOSE:

Before MR fingerprinting (MRF) can be adopted clinically, the derived quantitative values must be proven accurate and repeatable over a range of T1 and T2 values and temperatures. Correct assessment of accuracy and precision as well as comparison between measurements can only be performed when temperature is either controlled or corrected for. The purpose of this study was to investigate the temperature dependence of T1 and T2 MRF values and evaluate the accuracy and repeatability of temperature-corrected relaxation values derived from a B1 -corrected MRF-fast imaging with steady-state precession implementation using 2 different dictionary sizes.

METHODS:

The International Society of MR in Medicine/National Institute of Standards and Technology phantom was scanned using an MRF sequence of 2 different lengths, a variable flip angle T1 , and a multi-echo spin echo T2 at 14 temperatures ranging from 15°C to 28°C and investigated with a linear regression model. Temperature-corrected accuracy was evaluated by correlating T1 and T2 times from each MRF dictionary with reference values. Repeatability was assessed using the coefficient of variation, with measurements taken over 30 separate sessions.

RESULTS:

There was a statistically significant fit of the model for MRF-derived T1 and T2 and temperature (p < 0.05) for all the spheres with a T1 > 500 ms. Both MRF methods showed a strong linear correlation with reference values for T1 (R2 = 0.996) and T2 (R2 = 0.982). MRF repeatability for T1 values was ≤1.4% and for T2 values was ≤3.4%.

CONCLUSION:

MRF demonstrated relaxation times with a temperature dependence similar to that of conventional mapping methods. Temperature-corrected T1 and T2 values from both dictionaries showed adequate accuracy and excellent repeatability in this phantom study.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Brain / Magnetic Resonance Imaging Type of study: Guideline / Prognostic_studies Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Type: Article Affiliation country: United kingdom

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Brain / Magnetic Resonance Imaging Type of study: Guideline / Prognostic_studies Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2022 Type: Article Affiliation country: United kingdom