Your browser doesn't support javascript.
loading
Eddy current-induced artifact correction in high b-value ex vivo human brain diffusion MRI with dynamic field monitoring.
Ramos-Llordén, Gabriel; Park, Daniel J; Kirsch, John E; Scholz, Alina; Keil, Boris; Maffei, Chiara; Lee, Hong-Hsi; Bilgic, Berkin; Edlow, Brian L; Mekkaoui, Choukri; Yendiki, Anastasia; Witzel, Thomas; Huang, Susie Y.
Affiliation
  • Ramos-Llordén G; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Park DJ; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Kirsch JE; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Scholz A; Institute of Medical Physics and Radiation Protection, Mittelhessen University of Applied Sciences, Giessen, Germany.
  • Keil B; Institute of Medical Physics and Radiation Protection, Mittelhessen University of Applied Sciences, Giessen, Germany.
  • Maffei C; Department of Diagnostic and Interventional Radiology, University Hospital Marburg, Philipps University of Marburg, Marburg, Germany.
  • Lee HH; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Bilgic B; Center for Neurotechnology and Neurorecovery, Department of Neurology, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Edlow BL; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Mekkaoui C; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Yendiki A; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
  • Witzel T; Center for Neurotechnology and Neurorecovery, Department of Neurology, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Huang SY; Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital and Harvard Medical School, Charlestown, Massachusetts, USA.
Magn Reson Med ; 91(2): 541-557, 2024 Feb.
Article in En | MEDLINE | ID: mdl-37753621
ABSTRACT

PURPOSE:

To investigate whether spatiotemporal magnetic field monitoring can correct pronounced eddy current-induced artifacts incurred by strong diffusion-sensitizing gradients up to 300 mT/m used in high b-value diffusion-weighted (DW) EPI.

METHODS:

A dynamic field camera equipped with 16 1 H NMR field probes was first used to characterize field perturbations caused by residual eddy currents from diffusion gradients waveforms in a 3D multi-shot EPI sequence on a 3T Connectom scanner for different gradient strengths (up to 300 mT/m), diffusion directions, and shots. The efficacy of dynamic field monitoring-based image reconstruction was demonstrated on high-gradient strength, submillimeter resolution whole-brain ex vivo diffusion MRI. A 3D multi-shot image reconstruction framework was developed that incorporated the nonlinear phase evolution measured with the dynamic field camera.

RESULTS:

Phase perturbations in the readout induced by residual eddy currents from strong diffusion gradients are highly nonlinear in space and time, vary among diffusion directions, and interfere significantly with the image encoding gradients, changing the k-space trajectory. During the readout, phase modulations between odd and even EPI echoes become non-static and diffusion encoding direction-dependent. Superior reduction of ghosting and geometric distortion was achieved with dynamic field monitoring compared to ghosting reduction approaches such as navigator- and structured low-rank-based methods or MUSE followed by image-based distortion correction with the FSL tool "eddy."

CONCLUSION:

Strong eddy current artifacts characteristic of high-gradient strength DW-EPI can be well corrected with dynamic field monitoring-based image reconstruction.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Artifacts Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2024 Type: Article Affiliation country: United States

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Image Processing, Computer-Assisted / Artifacts Limits: Humans Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2024 Type: Article Affiliation country: United States