Human prostate MRI at ultrahigh-performance gradient: A feasibility study.
Magn Reson Med
; 91(2): 640-648, 2024 Feb.
Article
in En
| MEDLINE
| ID: mdl-37753628
ABSTRACT
PURPOSE:
To demonstrate the technical feasibility and the value of ultrahigh-performance gradient in imaging the prostate in a 3T MRI system.METHODS:
In this local institutional review board-approved study, prostate MRI was performed on 4 healthy men. Each subject was scanned in a prototype 3T MRI system with a 42-cm inner-diameter gradient coil that achieves a maximum gradient amplitude of 200 mT/m and slew rate of 500 T/m/s. PI-RADS V2.1-compliant axial T2 -weighted anatomical imaging and single-shot echo planar DWI at standard gradient of 70 mT/m and 150 T/m/s were obtained, followed by DWI at maximum performance (i.e., 200 mT/m and 500 T/m/s). In comparison to state-of-the-art clinical whole-body MRI systems, the high slew rate improved echo spacing from 1020 to 596 µs and, together with a high gradient amplitude for diffusion encoding, TE was reduced from 55 to 36 ms.RESULTS:
In all 4 subjects (waist circumference = 81-91 cm, age = 45-65 years), no peripheral nerve stimulation sensation was reported during DWI. Reduced image distortion in the posterior peripheral zone prostate gland and higher signal intensity, such as in the surrounding muscle of high-gradient DWI, were noted.CONCLUSION:
Human prostate MRI at simultaneously high gradient amplitude of 200 mT/m and slew rate of 500 T/m/s is feasible, demonstrating that improved gradient performance can address image distortion and T2 decay-induced SNR issues for in vivo prostate imaging.Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Prostatic Neoplasms
/
Magnetic Resonance Imaging
Limits:
Aged
/
Humans
/
Male
/
Middle aged
Language:
En
Journal:
Magn Reson Med
Journal subject:
DIAGNOSTICO POR IMAGEM
Year:
2024
Type:
Article
Affiliation country: