Your browser doesn't support javascript.
loading
Image-Guided Transbronchial Pulmonary Cryoablation with a Flexible Cryoprobe in Swine: Performance and Radiology-Pathology Correlation.
de Ruiter, Quirina M B; Mauda-Havakuk, Michal M; Starost, Matthew F; Bakhutashvili, Ivane; Esparza-Trujillo, Juan A; Brown, Andrew; Natesan, Harishankar; Kveen, Graig; Lewis, Andrew L; Wood, Bradford J; Pritchard, William F; Karanian, John W.
Affiliation
  • de Ruiter QMB; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland.
  • Mauda-Havakuk MM; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland; Interventional Radiology, Tel-Aviv Sourasky Medical Center, Tel-Aviv, Israel.
  • Starost MF; Division of Veterinary Resources, National Institutes of Health, Bethesda, Maryland.
  • Bakhutashvili I; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland.
  • Esparza-Trujillo JA; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland.
  • Brown A; Boston Scientific (formerly BTG), Arden Hills, Minnesota.
  • Natesan H; Boston Scientific (formerly BTG), Arden Hills, Minnesota.
  • Kveen G; Boston Scientific (formerly BTG), Arden Hills, Minnesota.
  • Lewis AL; Boston Scientific (formerly BTG), Arden Hills, Minnesota; Alchemed Bioscience Consulting Ltd, Farnham, Surrey, United Kingdom.
  • Wood BJ; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland; Center for Cancer Research, National Institutes of Health, Bethesda, Maryland.
  • Pritchard WF; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland.
  • Karanian JW; Center for Interventional Oncology, Radiology & Imaging Sciences, Clinical Center, National Institutes of Health, Bethesda, Maryland. Electronic address: john.karanian@nih.gov.
J Vasc Interv Radiol ; 35(7): 1022-1030.e4, 2024 Jul.
Article in En | MEDLINE | ID: mdl-38599280
ABSTRACT

PURPOSE:

To evaluate the performance of a prototype flexible transbronchial cryoprobe compared with that of percutaneous transthoracic cryoablation and to define cone-beam computed tomography (CT) imaging and pathology cryolesion features in an in vivo swine model. MATERIALS AND

METHODS:

Transbronchial cryoablation was performed with a prototype flexible cryoprobe (3 central and 3 peripheral lung ablations in 3 swine) and compared with transthoracic cryoablation performed with a commercially available rigid cryoprobe (2 peripheral lung ablations in 1 swine). Procedural time and cryoablation success rates for endobronchial navigation and cryoneedle deployment were measured. Intraoperative cone-beam CT imaging features of cryolesions were characterized and correlated with gross pathology and hematoxylin and eosin-stained sections of the explanted cryolesions.

RESULTS:

The flexible cryoprobe was successfully navigated and delivered to each target through a steerable guiding sheath (6/6). At 4 minutes after ablation, 5 of 6 transbronchial and 2 of 2 transthoracic cryolesions were visible on cone-beam CT. The volumes on cone-beam CT images were 55.5 cm3 (SE ± 8.0) for central transbronchial ablations (n = 2), 72.5 cm3 (SE ± 8.1) for peripheral transbronchial ablations (n = 3), and 79.5 cm3 (SE ±11.6) for peripheral transthoracic ablations (n = 2). Pneumothorax developed in 1 animal after transbronchial ablation and during ablation in the transthoracic cryoablation. Images of cryoablation zones on cone-beam CT correlated well with the matched gross pathology and histopathology sections of the cryolesions.

CONCLUSIONS:

Transbronchial cryoablation with a flexible cryoprobe, delivered through a steerable guiding sheath, is feasible. Transbronchial cryoablation zones are imageable with cone-beam CT, with gross pathology and histopathology similar to those of transthoracic cryoablation.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cryosurgery / Equipment Design / Cone-Beam Computed Tomography Limits: Animals Language: En Journal: J Vasc Interv Radiol / J. vasc. interv. radiol / Journal of vascular and interventional radiology Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2024 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cryosurgery / Equipment Design / Cone-Beam Computed Tomography Limits: Animals Language: En Journal: J Vasc Interv Radiol / J. vasc. interv. radiol / Journal of vascular and interventional radiology Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2024 Type: Article