Reversible Fusion-Fission MXene Fiber-Based Microelectrodes for Target-Specific Gram-Positive and Gram-Negative Bacterium Discrimination.
Anal Chem
; 96(23): 9317-9324, 2024 06 11.
Article
in En
| MEDLINE
| ID: mdl-38818541
ABSTRACT
Inaccurate or cumbersome clinical pathogen diagnosis between Gram-positive bacteria (G+) and Gram-negative (G-) bacteria lead to delayed clinical therapeutic interventions. Microelectrode-based electrochemical sensors exhibit the significant advantages of rapid response and minimal sample consumption, but the loading capacity and discrimination precision are weak. Herein, we develop reversible fusion-fission MXene-based fiber microelectrodes for G+/G- bacteria analysis. During the fissuring process, the spatial utilization, loading capacity, sensitivity, and selectivity of microelectrodes were maximized, and polymyxin B and vancomycin were assembled for G+/G- identification. The surface-tension-driven reversible fusion facilitated its reusability. A deep learning model was further applied for the electrochemical impedance spectroscopy (EIS) identification in diverse ratio concentrations of G+ and G- of (1100-1001) with higher accuracy (>93%) and gave predictable detection results for unknown samples. Meanwhile, the as-proposed sensing platform reached higher sensitivity toward E. coli (24.3 CFU/mL) and S. aureus (37.2 CFU/mL) in 20 min. The as-proposed platform provides valuable insights for bacterium discrimination and quantification.
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Microelectrodes
Language:
En
Journal:
Anal Chem
Year:
2024
Type:
Article
Affiliation country:
China