Your browser doesn't support javascript.
loading
In-line pressure-flow module for in vitro modelling of haemodynamics and biosensor validation.
Koenig, S C; Schaub, J D; Ewert, D L; Swope, R D.
Afiliación
  • Koenig SC; Physiology Research Branch, Clinical Sciences Division, Brooks AFB, TX 78235-5117, USA. sckoen01@homer.louisville.edu
Med Biol Eng Comput ; 35(5): 549-52, 1997 Sep.
Article en En | MEDLINE | ID: mdl-11536815
ABSTRACT
An in-line pressure-flow module for in vitro modelling of haemodynamics and biosensor validation has been developed. Studies show that good accuracy can be achieved in the measurement of pressure and of flow, in steady and pulstile flow systems. The model can be used for development, testing and evaluation of cardiovascular-mechanical-electrical anlogue models, cardiovascular prosthetics (i.e. valves, vascular grafts) and pressure and flow biosensors.
Asunto(s)
Palabras clave
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Transductores de Presión / Simulación por Computador / Técnicas Biosensibles / Hemodinámica / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Animals / Humans Idioma: En Revista: Med Biol Eng Comput Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Transductores de Presión / Simulación por Computador / Técnicas Biosensibles / Hemodinámica / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies Límite: Animals / Humans Idioma: En Revista: Med Biol Eng Comput Año: 1997 Tipo del documento: Article País de afiliación: Estados Unidos