[Effect of prolonging time for sampling on preservation of ultrastructure during the specimen preparation].
Jing, J H.
Afiliación
  • Jing JH; Laboratory of Electron Micrscopy, Hunan Medical University, Changsha 410078.
Hunan Yi Ke Da Xue Xue Bao ; 25(2): 205-6, 2000 Apr 28.
Article en Zh | MEDLINE | ID: mdl-12212227
The ultrastructural change of rat liver, which was not immediately fixed after separation from the animal, was observed under electron microscope. The results showed that the ultrastructure of hepatocytes still satisfied the needs for diagnostic electron microscopy if the isolated livers were soaked in normal saline for 30 minutes, at room temperature, or placed in 4 degrees C refrigerator for 60 minutes, then they were fixed. After the isolated samples were put at room temperature for over 30 minutes, ultrastructure of hepatocytes were damaged so severely as to lose value for observation under electron microscope.