Your browser doesn't support javascript.
loading
Simultaneous determination of spirapril and spiraprilat in plasma by capillary gas chromatography-mass spectrometry.
Schürer, Michael; Amschler, Stefan; Schulz, Hans-Ulrich; Schäfer, Harald F.
Afiliación
  • Schürer M; LAFAA Laboratory for Contract Research in Clinical Pharmacology and Biopharmaceutical Analytics GmbH, Bad Schwartau Germany. lafaa_gmbh@t-onbline.de
Arzneimittelforschung ; 53(5): 307-13, 2003.
Article en En | MEDLINE | ID: mdl-12854357
ABSTRACT
A specific, sensitive and precise method for the simultaneous determination of spirapril (CAS 94841-17-5) and spiraprilat (CAS 83602-05-5) in human plasma is described. The method involves the use of enalapril as internal standard, solid-phase extraction, derivatization and capillary gas chromatography with mass sensitive detection. The working range is from 2.5 to 500 micrograms/l for spirapril and spiraprilat, respectively. Data demonstrating the precision and accuracy of the analytical method are given. Moreover, data concerning freeze-thaw stability, long-term stability of frozen samples, short-term stability of thawed samples, and stability of the extracts in the autosampler are given.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Inhibidores de la Enzima Convertidora de Angiotensina / Enalapril Límite: Humans Idioma: En Revista: Arzneimittelforschung Año: 2003 Tipo del documento: Article
Buscar en Google