[Neonatal hearing loss screening: our results three years after starting the program]. / Despistaje de la hipoacusia neonatal: resultados después de 3 años de iniciar nuestro programa.
Acta Otorrinolaringol Esp
; 56(2): 55-8, 2005 Feb.
Article
en Es
| MEDLINE
| ID: mdl-15782642
Since the 90s there increasing interest in early detection of neonatal hearing loss due to the feasibility of universal neonatal screening as well as to the possibility of early intervention. This has led to set up neonatal screening programs based upon otoacoustic emissions and to a great increase in the number of publications in the field. Our hospital started a neonatal hearing loss screening program using otoacoustic emission in March 2000. We used a Capella Cochlear Emissions (MADSEN electronics) equipment. Until May 2003 we have performed the test in 6372 children with a median age of 71 days. In our study, the sensibility of the test was 90.9% and the specificity 99.7%. Our results are similar to those reported in the literature although the age of the children was higher and we have found an incidence of hearing loss lower than other studies.
Buscar en Google
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Tamizaje Neonatal
/
Pérdida Auditiva
Tipo de estudio:
Diagnostic_studies
/
Screening_studies
Límite:
Female
/
Humans
/
Infant
/
Male
/
Newborn
Idioma:
Es
Revista:
Acta Otorrinolaringol Esp
Año:
2005
Tipo del documento:
Article