The Vicenza ''Short'' peritoneal catheter: a twenty year experience.
Dell'aquila, R; Chiaramonte, S; Rodighiero, M P; Di Loreto, P; Spano', E; Nalesso, F; Cruz, D; Kuang, D; Ronco, C.
Afiliación
  • Dell'aquila R; Department of Nephrology, Dialysis and Transplantation, St. Bortolo Hospital, Vicenza, Italy. roberto.dellaquila@ulssvicenza.it
Int J Artif Organs ; 29(1): 123-7, 2006 Jan.
Article en En | MEDLINE | ID: mdl-16485247
Dislocation of peritoneal dialysis catheters is one of the major causes of technique failure. We evaluated 701 Vicenza catheters, implanted since 1985 in 365 males, mean age 53 +/- 16 yrs, range 24 - 87, and 336 females, mean age 51 +/- 17 yrs, range 21 - 82. The Vicenza catheter is defined "short" since it consists of a classic straight double cuff PD catheter having however an inner segment (the portion located in the peritoneal cavity) much shorter than any other type of catheter. It is implanted in the lower abdomen, just a few centimeters above the pubis. The analysis of our results obtained in a large PD population displayed good device survival at 2 and 5 years (94.3% and 91.5% respectively), a low dislocation rate (4%) and an exit-site infection rate similar to other double cuffed catheters. There was no selection of patients receiving this catheter since from 1985 we have used this catheter in every incident patient. Due to its lower implantation site this catheter demonstrates excellent wearability and good body image acceptance.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Catéteres de Permanencia / Diálisis Peritoneal Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Artif Organs Año: 2006 Tipo del documento: Article País de afiliación: Italia
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Catéteres de Permanencia / Diálisis Peritoneal Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Artif Organs Año: 2006 Tipo del documento: Article País de afiliación: Italia