A 4 x 500 mm2 cloverleaf detector system for in vivo bone lead measurement.
Med Phys
; 34(3): 945-51, 2007 Mar.
Article
en En
| MEDLINE
| ID: mdl-17441240
A 4 x 500 mm2 "cloverleaf" low energy germanium detector array has been assembled for the purpose of in vivo bone lead measurement through x-ray fluorescence. Using 109Cd as an exciting source, results are reported from a leg phantom simulating measurement of lead in a human tibia. For high activity (4.0-4.4 GBq) and low activity (0.18-0.19 GBq) sources, measurement results are reported for both the cloverleaf system and a conventional single detector system of equivalent surface area (2000 mm2). The mean uncertainty and reproducibility of measurement were both significantly improved for the cloverleaf system with a high activity 109Cd source. When using a source activity of 4.4 GBq, measurement of the phantom resulted in an average bone lead uncertainty of 0.79 microg/g and a reproducibility of 0.84 microg/g. These results represent the highest precision yet reported from a bone lead x-ray fluorescence system.
Buscar en Google
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Espectrometría por Rayos X
/
Huesos
/
Plomo
Tipo de estudio:
Diagnostic_studies
Límite:
Humans
Idioma:
En
Revista:
Med Phys
Año:
2007
Tipo del documento:
Article
País de afiliación:
Canadá