Entrance and exit TL-dosimetry in the conservative treatment of breast cancer: a pilot study for the EORTC-Radiotherapy Cooperative Group.
Radiother Oncol
; 22(4): 280-4, 1991 Dec.
Article
en En
| MEDLINE
| ID: mdl-1792321
Mailed dosimetry, using thermoluminescent dosimeters, can play an important role in quality assurance procedures in radiotherapy. In 1989, a pilot study was started with the main aim to show the feasibility of this method for the multicentre EORTC trial 22881 on the conservative management of breast carcinoma. Two anthropomorphic breast phantoms and six patients with breast carcinoma were irradiated according the prescriptions of the protocol. TLD measurements of the entrance and exit dose were performed in 6 MV tangential X-ray beams. It proved to be possible to correlate the dose measured in the entrance and exit points of the beams to the calculated dose closely under the surface. A thickness of at least 5 mm bolus material must be applied over the dosimeters and a distance of at least 3 cm from the lateral and medial field borders must be maintained in order to reach a clinically acceptable accuracy in the measurements.
Buscar en Google
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Dosificación Radioterapéutica
/
Dosimetría Termoluminiscente
/
Neoplasias de la Mama
Tipo de estudio:
Clinical_trials
/
Guideline
Límite:
Female
/
Humans
País/Región como asunto:
Europa
Idioma:
En
Revista:
Radiother Oncol
Año:
1991
Tipo del documento:
Article
País de afiliación:
Países Bajos