Your browser doesn't support javascript.
loading
The endotracheal tube biases the estimates of pulmonary recruitment and overdistension.
Jandre, Frederico C; Modesto, Felipe Cardozo; Carvalho, Alysson Roncally Silva; Giannella-Neto, Antonio.
Afiliación
  • Jandre FC; Biomedical Engineering Program, Federal University of Rio de Janeiro, Graduate School of Engineering, COPPE, Rio de Janeiro, Brazil.
Med Biol Eng Comput ; 46(1): 69-73, 2008 Jan.
Article en En | MEDLINE | ID: mdl-17932699
ABSTRACT
To assess the impact of the endotracheal tube (ETT) and of different flow waveforms on estimates of alveolar cyclic recruitment (CR) and overdistension (AO). Numerical simulation of the respiratory system plus ETT (inertance L plus a flow-dependent resistance, K (1) and K (2)), with the following non-linear equation of motion PAW(t)= ((K1 + K2 x/V(t)/) x V(t) + L x V(t)) + Rrs x V(t) + (E1 + E2 x V(t) x V(t) + P0 (PAW pressure at the airways opening, V volume), under volume-controlled mechanical ventilation. An index %E2 = 100 x (E2 x V(T))/(E1 + E2)x V(T)) can be calculated where %E(2) > 30% represents AO and %E(2) < 0% represents CR. Parameters were estimated by the least-squares method, either with the complete equation or suppressing L, K(2) or both. %E(2) is always underestimated (down to -152 percent points) with incomplete equations of motion. The estimation of %E (2) may be strongly biased in the presence of an ETT excluded from the estimation model.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Respiración Artificial / Intubación Intratraqueal / Modelos Biológicos Límite: Humans Idioma: En Revista: Med Biol Eng Comput Año: 2008 Tipo del documento: Article País de afiliación: Brasil
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Respiración Artificial / Intubación Intratraqueal / Modelos Biológicos Límite: Humans Idioma: En Revista: Med Biol Eng Comput Año: 2008 Tipo del documento: Article País de afiliación: Brasil