Dissipative cryogenic filters with zero dc resistance.
Bluhm, Hendrik; Moler, Kathryn A.
Afiliación
  • Bluhm H; Department of Physics, Stanford University, Stanford, California 94305, USA. tohbluhm@ fas.harvard.edu
Rev Sci Instrum ; 79(1): 014703, 2008 Jan.
Article en En | MEDLINE | ID: mdl-18248056
The authors designed, implemented, and tested cryogenic rf filters with zero dc resistance, based on wires with a superconducting core inside a resistive sheath. The superconducting core allows low frequency currents to pass with negligible dissipation. Signals above the cutoff frequency are dissipated in the resistive part due to their small skin depth. The filters consist of twisted wire pairs shielded with copper tape. Above approximately 1 GHz, the attenuation is exponential in omega, as typical for skin depth based rf filters. By using additional capacitors of 10 nF per line, an attenuation of at least 45 dB above 10 MHz can be obtained. Thus, one single filter stage kept at mixing chamber temperature in a dilution refrigerator is sufficient to attenuate room temperature black body radiation to levels corresponding to 10 mK above about 10 MHz.