Omron HEM-711 DLX home Blood pressure monitor passes the European Society of Hypertension International Validation Protocol.
Grim, Clarence E; Grim, Carlene M.
Afiliación
  • Grim CE; Medical College of Wisconsin and High Blood Pressure Consulting, Milwaukee, WI 53211, USA. lowerbp2@aol.com
Blood Press Monit ; 13(4): 225-6, 2008 Aug.
Article en En | MEDLINE | ID: mdl-18635978
The Omron HEM-711 DLX home Blood pressure monitor was tested using the European Society of Hypertension International Protocol in 33 patients. The average error was 0.9+/-5.2 (SD) (range: -12 to 10 mmHg)/-0.8+/-5.6 (-19 to 8). It can be recommended for use by patients. With all automatic devices it is, however, important to validate its accuracy in each individual patient who uses it.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitores de Presión Sanguínea / Monitoreo Ambulatorio de la Presión Arterial Tipo de estudio: Evaluation_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Blood Press Monit Asunto de la revista: ANGIOLOGIA Año: 2008 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitores de Presión Sanguínea / Monitoreo Ambulatorio de la Presión Arterial Tipo de estudio: Evaluation_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Blood Press Monit Asunto de la revista: ANGIOLOGIA Año: 2008 Tipo del documento: Article País de afiliación: Estados Unidos