Omron HEM-711 DLX home Blood pressure monitor passes the European Society of Hypertension International Validation Protocol.
Blood Press Monit
; 13(4): 225-6, 2008 Aug.
Article
en En
| MEDLINE
| ID: mdl-18635978
The Omron HEM-711 DLX home Blood pressure monitor was tested using the European Society of Hypertension International Protocol in 33 patients. The average error was 0.9+/-5.2 (SD) (range: -12 to 10 mmHg)/-0.8+/-5.6 (-19 to 8). It can be recommended for use by patients. With all automatic devices it is, however, important to validate its accuracy in each individual patient who uses it.