Comparison of refractive outcome using intraoperative biometry and partial coherence interferometry in silicone oil-filled eyes.
El-Baha, Samir Mohamed; Hemeida, Tarek Samir.
Afiliación
  • El-Baha SM; Department of Ophthalmology, Alexandria University, Alexandria, Egypt. dr_selbaha@yahoo.com
Retina ; 29(1): 64-8, 2009 Jan.
Article en En | MEDLINE | ID: mdl-18695632
ABSTRACT

INTRODUCTION:

A-scan ultrasound is currently the most widely used technique for axial length measurement; however, this method is not optimal in silicone-oil filled eyes. Two techniques that may be more accurate for axial length measurements in these eyes are intraoperative measurement of axial length after silicone oil removal or measurement by Laser interferometry using the IOL Master. The purpose of this study is to evaluate and compare the accuracy of both intraoperative biometry and partial coherence interferometry in silicone oil-filled eyes. PATIENT AND

METHODS:

Axial length measurement of 22 cataractous silicone-filled eyes of 21 patients using both IOL Master Biometry and intraoperative A-scan biometry was performed. IOL power was then computed using the SRK-T formula. Accuracy of intraoperative biometry and partial coherence interferometry was evaluated by determining the mean actual postoperative refractive error.

RESULTS:

Comparing the predictability of intraoperative A-scan biometry and IOL Master Biometry, the two techniques showed small predictive postoperative refractive errors without a statistically significant difference in the predictive errors of the two techniques.

CONCLUSION:

Both intraoperative biometry by A-scan ultrasonography and partial coherence interferometer by IOL Master proved to have good equal predictability for absolute postoperative refractive error in cataract surgery for eyes filled with silicone oil.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Errores de Refracción / Aceites de Silicona / Biometría / Interferometría Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Retina Año: 2009 Tipo del documento: Article País de afiliación: Egipto

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Errores de Refracción / Aceites de Silicona / Biometría / Interferometría Tipo de estudio: Prognostic_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Retina Año: 2009 Tipo del documento: Article País de afiliación: Egipto