Ophthalmic applicators: an overview of calibrations following the change to SI units.
Holmes, Shannon M; Micka, John A; DeWerd, Larry A.
Afiliación
  • Holmes SM; Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705, USA.
Med Phys ; 36(5): 1473-7, 2009 May.
Article en En | MEDLINE | ID: mdl-19544762
Since the NIST dose to water standard for 90Sr/90Y ophthalmic applicators was introduced, numerous sources have undergone calibration either at NIST or at the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL). From 1997 to 2008, 222 of these beta-emitting sources were calibrated at the UWADCL, and prior reference source strength values were available for 149 of these sources. A survey of UWADCL ophthalmic applicator calibrations is presented here, demonstrating an average discrepancy of -19% with a standard deviation of +/- 16% between prior reference values and the NIST-traceable UWADCL absorbed dose to water calibrations. Values ranged from -49% to +42%.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radiometría / Algoritmos / Braquiterapia / Neoplasias del Ojo / Sistema Internacional de Unidades Tipo de estudio: Diagnostic_studies Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: Med Phys Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Radiometría / Algoritmos / Braquiterapia / Neoplasias del Ojo / Sistema Internacional de Unidades Tipo de estudio: Diagnostic_studies Límite: Humans País/Región como asunto: America do norte Idioma: En Revista: Med Phys Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos