A protocol design for evaluation of wearable cuff-less blood pressure measuring devices.
Annu Int Conf IEEE Eng Med Biol Soc
; 2009: 7045-7, 2009.
Article
en En
| MEDLINE
| ID: mdl-19964197
ABSTRACT
This paper proposes one of the unique requirements in the validation protocol of the IEEE P1708 standard in assessing wearable cuff-less blood pressure (BP) measuring devices. Based on principles that are different from that of the conventional cuff-based devices, the cuff-less BP measurement approaches often require an individual calibration procedure. In this study, we used data from an experiment carried out on 28 subjects with a total of 139 sets of BP measurements as an example to show that breakdown of the performance evaluation of cuff-less devices according to the change of BP from the point of calibration is crucial for understanding and interpreting the overall accuracy of the device.