A protocol design for evaluation of wearable cuff-less blood pressure measuring devices.
Yan, Iris R F; Poon, Carmen C Y; Zhang, Y T.
Afiliación
  • Yan IR; Joint Research Center for Biomedical Engineering (JCBME), Dept. of Electronic Engineering, The Chinese University of Hong Kong, Hong Kong, China.
Article en En | MEDLINE | ID: mdl-19964197
ABSTRACT
This paper proposes one of the unique requirements in the validation protocol of the IEEE P1708 standard in assessing wearable cuff-less blood pressure (BP) measuring devices. Based on principles that are different from that of the conventional cuff-based devices, the cuff-less BP measurement approaches often require an individual calibration procedure. In this study, we used data from an experiment carried out on 28 subjects with a total of 139 sets of BP measurements as an example to show that breakdown of the performance evaluation of cuff-less devices according to the change of BP from the point of calibration is crucial for understanding and interpreting the overall accuracy of the device.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial Tipo de estudio: Evaluation_studies / Guideline Límite: Adult / Humans Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Año: 2009 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Monitoreo Ambulatorio de la Presión Arterial Tipo de estudio: Evaluation_studies / Guideline Límite: Adult / Humans Idioma: En Revista: Annu Int Conf IEEE Eng Med Biol Soc Año: 2009 Tipo del documento: Article País de afiliación: China