Performance of a new blood glucose monitoring system in the hands of intended users.
Bailey, Timothy S; Wallace, Jane F; Parkes, Joan Lee; Pardo, Scott; Yu, Ji; Schachner, Holly C; Simmons, David A; Chu, Amy.
Afiliación
  • Bailey TS; AMCR Institute Inc, Escondido, California 92026, USA. tbailey@amcrinstitute.com
Diabetes Technol Ther ; 14(9): 783-9, 2012 Sep.
Article en En | MEDLINE | ID: mdl-22746838
ABSTRACT

BACKGROUND:

This study assessed the performance of a blood glucose monitoring system (BGMS) in development that uses a new generation of blood glucose test strips with capillary and venous blood in the hands of its intended users, people with diabetes and healthcare professionals (HCPs). SUBJECTS AND

METHODS:

In total, 93 subjects ≥ 18 years old (median age, 33 years) with type 1 (78%) or type 2 (22%) diabetes participated. Untrained subjects performed self-test fingersticks using a Microlet(®)2 lancing device (Bayer HealthCare LLC, Diabetes Care, Tarrytown, NY) followed by testing of their own capillary blood on the BGMS. HCPs performed fingersticks (using a Tenderlett(®) lancing device [International Technidyne Corp., Edison, NJ]) and venipunctures on subjects and tested both capillary and venous samples from subjects on the BGMS. All BGMS results were compared with Yellow Springs Instruments (YSI) (YSI Life Sciences, Inc., Yellow Springs, OH) laboratory results. Analytical accuracy was assessed according to International Organization for Standardization (ISO) 151972003 guidelines (i.e., within ± 15 mg/dL or ± 20% of the YSI results for samples with glucose concentrations < 75 mg/dL and ≥ 75 mg/dL, respectively) and more stringent criteria (i.e., within ± 15 mg/dL or ± 15% of the YSI results for samples with glucose concentrations < 100 mg/dL and ≥ 100 mg/dL, respectively).

RESULTS:

Overall, 98.9% (180/182) of subject Microlet2 capillary fingerstick results, 99.5% (182/183) of HCP Tenderlett capillary fingerstick results, and 100% (186/186) of venous results met current ISO criteria and more stringent criteria. The average hematocrit was 44%, with values ranging from 32% to 52%.

CONCLUSIONS:

Test results from both capillary fingerstick and venous samples with a new BGMS in development met current accuracy guidelines as well as proposed tighter criteria.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Glucemia / Automonitorización de la Glucosa Sanguínea / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 / Hematócrito Tipo de estudio: Evaluation_studies / Qualitative_research Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: Diabetes Technol Ther Asunto de la revista: ENDOCRINOLOGIA / TERAPEUTICA Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Glucemia / Automonitorización de la Glucosa Sanguínea / Diabetes Mellitus Tipo 1 / Diabetes Mellitus Tipo 2 / Hematócrito Tipo de estudio: Evaluation_studies / Qualitative_research Límite: Adolescent / Adult / Aged / Female / Humans / Male / Middle aged País/Región como asunto: America do norte Idioma: En Revista: Diabetes Technol Ther Asunto de la revista: ENDOCRINOLOGIA / TERAPEUTICA Año: 2012 Tipo del documento: Article País de afiliación: Estados Unidos